SU-F-T-311: Comparison of Measurements From Two Diode Array QA Devices to Deep Point Dose Measurements with Two Treatment Planning Model Settings for Brain VMAT SRT Patient-Specific QA.

Medical Physics 2016 June
PURPOSE: To investigate the sensitivity of traditional gamma-index-based analysis using diode arrays to detect significant dosimetric errors in patientspecific measurements for VMAT brain SRT plans.

METHODS: Quality assurance measurements were performed using an ion chamber and two different diode array devices for four brain SRT treatment plans. All cases were planned with Eclipse AAA RapidArc using 6MV VMAT for a TrueBeam STx. Ion chambers measurements were performed with a PinPoint 3D micro-ion chamber in both the Standard Imaging Stereotactic Dose Verification (SI) Phantom and in the ArcCheck (AC) cavity plug. Point dose measurements were taken at isocenter within the target. Array measurements were acquired with the ArcCheck and gantry-mounted MapCheck (MC) and assessed using the SNC patient software with a 3%/3mm absolute dose gamma-index analysis with global normalization and a 10% dose threshold. Calculations and their corresponding measurements were performed and compared before and after a dynamic leaf gap (DLG) adjustment.

RESULTS: The measured doses at isocenter dropped from 6.3%-8.7% in the SI phantom and 4.0%-9.3% for AC down to -0.1%-1.1% in the SI phantom and 0.8%-2.9% in the AC after adjusting the DLG. However, the cumulative dose differences to MC's central diode ranged -2.0%-1.9% before the DLG adjustment and were -7.7%-0.2% afterwards. Similarly, the gamma-index analysis results for MC were found to be 94.1%-100% before the DLG adjustment and 94.4%-100% afterwards. The gamma-index results from AC were 72.9%-98.7% before and 95.9%-99.7% after.

CONCLUSION: Our data suggest that traditional passing criteria using gantry-mounted MapCheck may not have the sensitivity to catch significant dose errors introduced by DLG settings. ArcCheck measurements did flag some dose differences, however there appears to be a possibility that dose discrepancies may not always be observed for VMAT SRT treatments, and point dose measurements are recommended for verification.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app