Free-breathing dynamic liver examination using a radial 3D T1-weighted gradient echo sequence with moderate undersampling for patients with limited breath-holding capacity.

PURPOSE: To compare free-breathing radial VIBE with moderate undersampling (us-radial-VIBE) with a standard breathhold T1-weighted volumetric interpolated sequence (3D GRE VIBE) in patients unable to suspend respiration during dynamic liver examination.

MATERIAL AND METHODS: 23 consecutive patients underwent dynamic liver MR examination using the free-breathing us-radial-VIBE sequence as part of their oncologic follow-up. All patients were eligible for the free-breathing protocol due to severe respiratory artifacts at the planning or precontrast sequences. The us-radial-VIBE acquisitions were compared to the patientś last staging liver MRI including a standard breathhold 3D GRE VIBE. For an objective image evaluation, signal intensity (SI), image noise (IN), signal-to-noise ratio (SNR) and contrast-enhancement ratio (CER) were compared. Representative image quality parameters, including typical artifacts were independently, retrospectively and blindly scored by four readers.

RESULTS: Us-radial-VIBE had significant lower SNR (p<0.0001) and higher IN (p<0.0001), whereas SI did not differ (p=0.62). Temporal resolution assessed with CER in the arterial phase showed higher values for us-radial-VIBE (p=0.028). Subjective image quality parameters received generally slightly higher scores for 3D GRE VIBE. In a smaller subgroup comprising patients with severe respiratory artifacts also at reference breathhold 3D GRE VIBE examination, us-radial-VIBE showed significantly higher image quality scores. Furthermore, there were generally more severe respiratory artifacts in 3D GRE VIBE, whereas streaking was characteristic in almost all us-radial-VIBE acquisitions but did not affect diagnostic validity.

CONCLUSION: Free-breathing dynamic liver imaging using us-radial-VIBE delivers accurate temporal resolution, low motion artifact susceptibility and good image quality and represents a promising alternative in patients unable to suspend respiration.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app