Add like
Add dislike
Add to saved papers

An Innovative Air-Oxygen Blender for Continuous Positive Airway Pressure Support in Resource-Poor Locations: A Feasibility Study.

Objectives: The objectives of this study are (i) to evaluate the feasibility of using an inexpensive air pump to maintain reliable oxygen concentration in a continuous positive airway pressure (CPAP) system and (ii) to evaluate whether an inexpensive air pump can maintain infant 02 sats >90%.

Methods: This prospective study, which included 19 babies in pilot phase and 90 during extension phase, was conducted at a neonatal intensive care unit in a resource-poor academic medical center in India. The intervention involved introduction of an air pump in the CPAP delivery system. Outcome measures were oxygen concentration in the air-oxygen blend and oxygen saturation of the study babies.

Results: Oxygen concentration at the outlets ranged between 56 and 70% and in the blend between 42 and 51%. Oxygen saturation ranged between 90 and 97%.

Conclusion: A simple and inexpensive air pump can work as a safe and effective oxygen blender.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app