Long-Term Valve Performance of TAVR and SAVR: A Report From the PARTNER I Trial.

OBJECTIVES: The aim of this study was to evaluate the long-term performance of transcatheter aortic valve replacement (TAVR) and surgical aortic valve replacement (SAVR) through longitudinal echocardiographic analysis.

BACKGROUND: The long-term performance of the SAPIEN TAVR is not well-described. Therefore, we examined the hemodynamic and valvular profile of the SAPIEN TAVR over 5 years.

METHODS: All patients receiving TAVR or SAVR with first post-implant (FPI) and 5-year echoes were analyzed for aortic valve (AV) peak velocity, AV mean gradient, AV area, peak left ventricular (LV) outflow tract and in-stent velocities, Doppler velocity index, aortic regurgitation (AR), LV mass index, stroke volume index, and cardiac index. The FPI and 5-year data were compared using a paired t test or McNemar's analyses.

RESULTS: There were 86 TAVR and 48 SAVR patients with paired FPI and 5-year echocardiograms. Baseline characteristics were similar between groups. The AV area did not change significantly 5 years after TAVR (p = 0.35). The AV mean gradient also remained stable: 11.5 ± 5.4 mm Hg at FPI to 11.0 ± 6.3 mm Hg at 5 years (p = 0.41). In contrast, the peak AV and LV outflow tract velocities decreased (p = 0.03 and p = 0.008, respectively), as did in-stent velocity (p = 0.015). Correspondingly, the TAVR Doppler velocity index was unchanged (p = 0.07). Among TAVR patients, there was no change in total AR (p = 0.40), transvalvular AR (p = 0.37), or paravalvular AR (p = 0.26). Stroke volume index and cardiac index remained stable (p = 0.16 and p = 0.25, respectively). However, there was a significant regression of LV mass index (p < 0.0001). The longitudinal evaluation among SAVR patients revealed similar trends. There was a low rate of adverse events among TAVR and SAVR patients alive at 5 years.

CONCLUSIONS: Longitudinal assessment of the PARTNER I trial (THE PARTNER TRIAL: Placement of AoRTic TraNscathetER Valve Trial) demonstrates that valve performance and cardiac hemodynamics are stable after implantation in both SAPIEN TAVR and SAVR in patients alive at 5 years. (THE PARTNER TRIAL: Placement of AoRTic TraNscathetER Valve Trial; NCT00530894).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app