All-Suture Anchors: Biomechanical Analysis of Pullout Strength, Displacement, and Failure Mode.

Arthroscopy 2017 June
PURPOSE: To evaluate the biomechanical and design characteristics of all-suture anchors.

METHODS: All-suture anchors were tested in fresh porcine cortical bone and biphasic polyurethane foam blocks by cyclic loading (10-100 N for 200 cycles), followed by destructive testing parallel to the insertion axis at 12.5 mm/s. Endpoints included ultimate failure load, displacement at 100 and 200 cycles, stiffness, and failure mode. Anchors tested included JuggerKnot (1.4, 1.5, and 2.8), Iconix (1, 2, and 3), Y-knot (1.3, 1.8, and 2.8), Q-Fix (1.8 and 2.8), and Draw Tight (1.8 and 3.2).

RESULTS: The mean ultimate failure strength of the triple-loaded anchors (564 ± 42 N) was significantly greater than the mean ultimate failure strength of the double-loaded anchors (465 ± 33 N) (P = .017), and the double-loaded anchors were stronger than the single-loaded anchors (256 ± 35 N) (P < .0001). No difference was found between the results in porcine bone and biphasic polyurethane foam. None of these anchors demonstrated 5 mm or 10 mm of displacement during cyclic loading. The Y-Knot demonstrated greater displacement than the JuggerKnot and Q-Fix (P = .025) but not the Iconix and Draw Tight (P > .05). The most common failure mode varied and was suture breaking for the Q-Fix (97%), JuggerKnot (81%), and Iconix anchors (58%), anchor pullout with the Draw Tight (76%), whereas the Y-Knot was 50% suture breaking and 50% anchor pullout.

CONCLUSIONS: The ultimate failure load of an all-suture anchor is correlated directly with its number of sutures. With cyclic loading, the Y-Knot demonstrated greater displacement than the JuggerKnot and Q-Fix but not the Iconix and Draw Tight. JuggerKnot (81%) and Q-Fix (97%) anchors failed by suture breaking, whereas the Draw Tight anchor failed by anchor pullout (76%).

CLINICAL RELEVANCE: All-suture anchors vary in strength and performance, and these factors may influence clinical success. Biphasic polyurethane foam is a validated model for suture anchor testing.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app