Redesign of an Open-System Oxygen Face Mask With Mainstream Capnometer for Children.

Respiratory Care 2017 January
BACKGROUND: Partial pressure of end-tidal carbon dioxide (PETCO2 ) monitoring in children is important to detect apnea or hypopnea early to intervene before hypoxemia develops. Monitoring PETCO2 in children without a tracheal tube is challenging. To improve PETCO2 measurement accuracy in a commercially available mask with a mainstream CO2 detector, we implemented design changes with deform-and-hold shaping technology and anterior-posterior adjustment of the expiratory gas flow cup.

METHODS: Two sizes of redesigned face masks (small for 7-20 kg, medium for 10-40 kg) were evaluated. Initial bench testing used a simulator modeling a spontaneously breathing infant and child with a natural airway. An infant/child manikin head was connected to the breathing lung simulator. A mass flow controller provided expiratory CO2. Mask fit was then evaluated on healthy human subjects to identify anatomical features associated with good fit, defined as square shape capnography waveform during expiration. A 3-dimensional digital scan was used to quantify anatomical features. The gaps between face mask rims and facial surface were manually measured.

RESULTS: Bench testing revealed a PETCO2 difference of 3.4 ± 1.5 mm Hg between a measured PETCO2 by the redesigned mask and CO2 concentration at trachea, as compared with 6.7 ± 6.2 mm Hg between PETCO2 measured by nasal cannula and trachea (P < .001). In the human mask fit study, 35 children (13 ± 4 kg) with the small mask and 38 (24 ± 8 kg) with the medium mask were evaluated. Capnography tracing was successfully obtained in 86% of the small and 100% of the medium masks. In children with small-size masks, the gap between the face mask rim and the child's face was not statistically different among those with good mask fit and without (1.0 ± 1.5 mm vs 1.4 ± 1.9 mm, P = .73).

CONCLUSIONS: PETCO2 measurement by a redesigned open-system face mask with a mainstream CO2 detector was accurate in the bench setting. The redesigned face mask can attain good mask fit and accurate capnography tracings in the majority of infants and children.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app