Evaluation of a Powered Stapler System with Gripping Surface Technology on Surgical Interventions Required During Laparoscopic Sleeve Gastrectomy.
BACKGROUND: Transection of gastric tissue during laparoscopic sleeve gastrectomy (LSG) can be challenging. Reinforcing the staple line may decrease the incidence of issues requiring intervention.
METHODS: The objective of this study was to compare the number of intraoperative surgical interventions for a surgical stapler and reload system with Gripping Surface Technology (GST) to standard reloads in patients who underwent LSG. Patients who underwent elective LSG were enrolled. The study was conducted in two stages. For Stage 1, procedures were performed using a powered stapler and standard reloads. For Stage 2, a reload system with GST was used. The primary endpoint was surgical interventions for bleeding and/or staple line issues during transection of the greater curvature of the stomach. Propensity score matching was applied to create two groups similar in baseline characteristics and risk factors.
RESULTS: A total of 111 subjects were enrolled across four centers. Propensity-matched procedures were completed with the standard (n = 38) or GST reloads (n = 38). The mean number of interventions in the standard group was 1.9 (1.29) versus 1.1 (1.45) in the GST group. Nonparametric comparisons were statistically significant, indicating a reduction in the distribution of interventions for GST subjects (P = .0036 for matched pair data). Tissue slippage during transection was low for both groups. Intraoperative leak testing was negative in all procedures, and no procedures were converted to open.
CONCLUSIONS: Use of the GST stapling system reduces the need for staple line interventions in LSG. Both stapling systems had an acceptable safety profile.
METHODS: The objective of this study was to compare the number of intraoperative surgical interventions for a surgical stapler and reload system with Gripping Surface Technology (GST) to standard reloads in patients who underwent LSG. Patients who underwent elective LSG were enrolled. The study was conducted in two stages. For Stage 1, procedures were performed using a powered stapler and standard reloads. For Stage 2, a reload system with GST was used. The primary endpoint was surgical interventions for bleeding and/or staple line issues during transection of the greater curvature of the stomach. Propensity score matching was applied to create two groups similar in baseline characteristics and risk factors.
RESULTS: A total of 111 subjects were enrolled across four centers. Propensity-matched procedures were completed with the standard (n = 38) or GST reloads (n = 38). The mean number of interventions in the standard group was 1.9 (1.29) versus 1.1 (1.45) in the GST group. Nonparametric comparisons were statistically significant, indicating a reduction in the distribution of interventions for GST subjects (P = .0036 for matched pair data). Tissue slippage during transection was low for both groups. Intraoperative leak testing was negative in all procedures, and no procedures were converted to open.
CONCLUSIONS: Use of the GST stapling system reduces the need for staple line interventions in LSG. Both stapling systems had an acceptable safety profile.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app