[Three successive fractures of different hip femoral stems on the same patient].
Acta Ortopédica Mexicana 2016 May
The clinical case of an 80-year-old female who suffered three fractures in uncemented stems implanted in her right hip in a period of three years is presented. A fracture occurred in the prosthetic neck and the other two at the juncture of the conical and cylindrical stem portion, coinciding with the metaphyseal-diaphyseal junction of the femur. The main causes of the failure were an increasing concentration of forces at the level of the implant as a consequence of the increased length of the neck and lateralization of the femur (offset). Other causes that have contributed to this prosthetic failure are analyzed.