ENGLISH ABSTRACT
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

[Three successive fractures of different hip femoral stems on the same patient].

The clinical case of an 80-year-old female who suffered three fractures in uncemented stems implanted in her right hip in a period of three years is presented. A fracture occurred in the prosthetic neck and the other two at the juncture of the conical and cylindrical stem portion, coinciding with the metaphyseal-diaphyseal junction of the femur. The main causes of the failure were an increasing concentration of forces at the level of the implant as a consequence of the increased length of the neck and lateralization of the femur (offset). Other causes that have contributed to this prosthetic failure are analyzed.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app