A Prospective Randomized Study Comparing Disposable with Reusable Blades for a Morcellator Device.

INTRODUCTION & OBJECTIVES: Transurethral enucleation of the prostate for the management of benign prostatic obstruction (BPO) involves two steps: the enucleation and morcellation procedure. The aim of our study was to assess the efficacy of a morcellator device using disposable and reusable blades with different settings of morcellation speed.

METHODS: A prospective randomized study was initiated for patients with symptomatic BPO undergoing Thulium laser enucleation of the prostate. Mechanical tissue morcellation was performed using the Piranha™ morcellator (R. Wolf, Knittlingen, Germany) with disposable or reusable blades at 850 (n = 24) or 1500 revolutions per minute (rpm) (n = 24). Patient characteristics, intraoperative complications, and the morcellation rate (g/min) were recorded. Data are expressed as median and interquartile range (IQR).

RESULTS: Forty-eight patients were randomized using disposable (n = 24) or reusable blades (n = 24). For reusable blades, the morcellation rate did not increase when changing the morcellation speed from 850 to 1500 rpm (5 vs 4.53 g/min, p = 0.843). The morcellation rate increased significantly when changing the morcellation speed from 850 to 1500 rpm using single-use blades (4.77 vs 10 g/min, p ≤ 0.014). The morcellation rate was not different at 850 rpm between reusable and single-use blades (5 vs 4.77 g/min, p = 0.671). Conversely, the morcellation rate was significantly different at 1500 rpm between reusable and single-use blades (4.53 vs 10 g/min, p ≤ 0.017). The total morcellation rate (at 850 and 1500 rpm) was significantly increased using single-use blades compared to reusable blades (7.67 vs 4.8 g/min, p ≤ 0.026). Interestingly, enucleated weight (g) and the morcellation rate (g/min) correlated inversely using single-use blades at 1500 rpm (r = -0.742, p ≤ 0.004). Only one superficial bladder injury occurred at 1500 rpm, which needed no further interventions.

CONCLUSIONS: The Piranha morcellator facilitates efficient tissue removal with single-use and reusable blades. Disposable morcellator blades increase tissue removal significantly at 1500 rpm.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app