Automated Measurement of Tear Film Dynamics and Lipid Layer Thickness for Assessment of Non-Sjögren Dry Eye Syndrome With Meibomian Gland Dysfunction.

Cornea 2016 December 9
PURPOSE: To investigate automated values from an advanced corneal topographer with a built-in real keratometer, color camera, and ocular surface interferometer for the evaluation of non-Sjögren dry eye syndrome (NSDES) with meibomian gland dysfunction (MGD).

METHODS: Sixty-four patients (64 eyes) diagnosed with NSDES with MGD were enrolled. All eyes were evaluated using the Ocular Surface Disease Index (OSDI), fluorescence staining score, tear film breakup time (TBUT), Schirmer test, and MGD grade. Noninvasive Keratograph average tear film breakup time (NIKBUTav), tear meniscus height (TMHk), meibomian gland (MG) dropout grade, and lipid layer thickness (LLT) using interferometry were measured.

RESULTS: Among automated indexes, NIKBUTav (mean 7.68 ± 4.07 s) and the MG dropout grade (mean 1.0 ± 0.5) significantly correlated with the OSDI (mean 40.6 ± 22.9) (r = -0.337, P = 0.006; and r = 0.201, P = 0.023, respectively), as did all conventional indicators, except the Schirmer score (mean 9.1 ± 5.9 mm). TMHk (mean 0.21 ± 0.18 mm) had significant correlation with the Schirmer score, the staining score (mean 1.2 ± 0.7), TBUT (mean 3.8 ± 1.8 s), and NIKBUTav (r = 0.298, P = 0.007; r = -0.268, P = 0.016; r = 0.459, P < 0.001; and r = 0.439, P < 0.001, respectively), but not any MGD indicator, even the MG dropout grade. NIKBUTav showed significant correlations with all clinical parameters and other automated values, except the Schirmer score and LLT (mean 83.94 ± 20.82 nm) (all |r| ≥ 0.25 and P < 0.01). The MG dropout grade highly correlated with all indexes except TMHk (all |r| ≥ 0.25 and P < 0.05). LLT was significantly associated with TBUT, MGD grade (mean 2.0 ± 0.7), and MG dropout grade (r = 0.219, P = 0.047; r = -0.221, P = 0.039; and r = 0.433, P < 0.001, respectively), although it was not related to patient symptoms.

CONCLUSIONS: Automated noninvasive measurements using an advanced corneal topographer and LLT measured with an ocular surface interferometer can be alternatives to conventional methods to evaluate tear conditions on the ocular surface; the former device can provide information about conformational MG changes in NSDES with MGD.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app