The association between defibrillation shock energy and acute cardiac damage in patients with implantable cardioverter defibrillators.
Journal of Arrhythmia 2016 December
BACKGROUND: The aim of this study was to establish a minimally invasive defibrillation testing (DT) protocol for patients with implantable cardioverter defibrillators (ICDs).
METHODS: Two different energy DTs were performed, immediately after (15 J-DT) and 7 days after (≤10 J-DT) device implantation, in 20 consecutive ICD implantation patients. Cardiac-troponin T (c-TNT) and heart-type fatty acid binding protein (H-FABP) levels were measured before implantation, 2 h after implantation, and 1 day after each DT. For an additional 122 patients with ICD, we retrospectively analyzed 203 DTs immediately and 7 days after device implantation.
RESULTS: Serum c-TNT levels were significantly elevated 2 h after 15 J-DT [0.008 (0.004-0.019) vs. 0.053 (0.037-0.068) ng/mL, p<0.001], but not ≤10 J-DT [0.007 (0.004-0.018) ng/mL]. Similarly, serum H-FABP levels were significantly elevated 2 h after 15 J-DT (2.9±1.5 vs. 6.4±3.4 ng/mL, p<0.001), but not ≤10 J-DT (2.7±1.5 ng/mL). The changes in c-TNT and H-FABP levels between baseline and 2 h after DT were significantly greater for 15 J-DT compared with ≤10 J-DT [c-TnT: 0.039 (0.029-0.060) vs. 0 (0-0.003) ng/mL, p<0.001; H-FABP: 3.6±2.8 vs. -0.16±1.1 ng/mL, p<0.001]. The success rates of the initial shocks delivered for ventricular fibrillation were no different between ≤10 J-DT (85% [78/92]) and ≥15 J-DT (92% [103/111]).
CONCLUSIONS: Elevated levels of myocardial damage markers such as c-TNT and H-FABP were not found after ≤10 J-DT. In addition, an acceptable success rate was confirmed in ≤10 J-DT.
METHODS: Two different energy DTs were performed, immediately after (15 J-DT) and 7 days after (≤10 J-DT) device implantation, in 20 consecutive ICD implantation patients. Cardiac-troponin T (c-TNT) and heart-type fatty acid binding protein (H-FABP) levels were measured before implantation, 2 h after implantation, and 1 day after each DT. For an additional 122 patients with ICD, we retrospectively analyzed 203 DTs immediately and 7 days after device implantation.
RESULTS: Serum c-TNT levels were significantly elevated 2 h after 15 J-DT [0.008 (0.004-0.019) vs. 0.053 (0.037-0.068) ng/mL, p<0.001], but not ≤10 J-DT [0.007 (0.004-0.018) ng/mL]. Similarly, serum H-FABP levels were significantly elevated 2 h after 15 J-DT (2.9±1.5 vs. 6.4±3.4 ng/mL, p<0.001), but not ≤10 J-DT (2.7±1.5 ng/mL). The changes in c-TNT and H-FABP levels between baseline and 2 h after DT were significantly greater for 15 J-DT compared with ≤10 J-DT [c-TnT: 0.039 (0.029-0.060) vs. 0 (0-0.003) ng/mL, p<0.001; H-FABP: 3.6±2.8 vs. -0.16±1.1 ng/mL, p<0.001]. The success rates of the initial shocks delivered for ventricular fibrillation were no different between ≤10 J-DT (85% [78/92]) and ≥15 J-DT (92% [103/111]).
CONCLUSIONS: Elevated levels of myocardial damage markers such as c-TNT and H-FABP were not found after ≤10 J-DT. In addition, an acceptable success rate was confirmed in ≤10 J-DT.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app