Finding the right fit: studying the biomechanics of under-tapping with varying thread depths and pitches.

BACKGROUND CONTEXT: Compromise of pedicle screw purchase is a concern in maintaining rigid spinal fixation, especially with osteoporosis. Little consistency exists among various tapping techniques. Pedicle screws are often prepared with taps of a smaller diameter, which can further exacerbate inconsistency.

PURPOSE: The objective of this study was to determine whether a mismatch between tap thread depth (D) and thread pitch (P) and screw D and P affects fixation when under-tapping in osteoporotic bone.

STUDY DESIGN: This study is a polyurethane foam block biomechanical analysis.

MATERIALS AND METHODS: A foam block osteoporotic bone model was used to compare pullout strength of pedicle screws with a 5.3 nominal diameter tap of varying D's and P's. Blocks were sorted into seven groups: (1) probe only; (2) 0.5-mm D, 1.5-mm P tap; (3) 0.5-mm D, 2.0-mm P tap; (4) 0.75-mm D, 2.0-mm P tap; (5) 0.75-mm D, 2.5-mm P tap; (6) 0.75-mm D, 3.0-mm P tap; and (7) 1.0-mm D, 2.5-mm P tap. A pedicle screw, 6.5 mm in diameter and 40 mm in length, was inserted to a depth of 40 mm. Axial pullout testing was performed at a rate of 5 mm/min on 10 blocks from each group.

RESULTS: No significant difference was noted between groups under axial pullout testing. The mode of failure in the probe-only group was block fracture, occurring in 50% of cases. Among the other six groups, only one screw failed because of block fracture. The other 59 failed because of screw pullout.

CONCLUSIONS: In an osteoporotic bone model, changing the D or P of the tap has no statistically significant effect on axial pullout. Osteoporotic bone might render tap features marginal. Our findings indicate that changing the characteristics of the tap D and P does not help with pullout strength in an osteoporotic model. The high rate of fracture in the probe-only group might imply the potential benefit of tapping to prevent catastrophic failure of bone.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app