Dosimetric characterization of MRI-only treatment planning for brain tumors in atlas-based pseudo-CT images generated from standard T1-weighted MR images.

Medical Physics 2016 December
PURPOSE: Magnetic resonance imaging (MRI)-only radiotherapy treatment planning requires accurate pseudo-CT (pCT) images for precise dose calculation. The current work introduced an atlas-based method combined with MR intensity information. pCT analyses and Monte Carlo dose calculations for intracranial stereotactic treatments were performed.

METHODS: Twenty-two patients, representing 35 tumor targets, were scanned using a 3D T1-weighted MRI sequence according to the clinical protocol. The MR atlas image was registered to the MR patient image using a deformable algorithm, and the deformation was then applied to the atlas CT. Two methods were applied. The first method (MRdef) was based on deformations only, while the second (MRint) also used the actual MR intensities. pCT analysis was performed using the mean (absolute) error, as well as an in-house tool based on a gamma index. Dose differences between pCT and true CT were analyzed using dose-volume histogram (DVH) parameters, statistical tests, the gamma index, and probability density functions. An unusual case, where the patient underwent an operation (part of the skull bone was removed), was studied in detail.

RESULTS: Soft tissues presented a mean error inferior to 50 HUs, while low-density tissues and bones presented discrepancies up to 600 HUs for hard bone. The MRdef method led to significant dose differences compared with the true CT (p-value < 0.05; Wilcoxon-signed-rank test). The MRint method performed better. The DVH parameter differences compared with CT were between -2.9% and 3.1%, except for two cases where the tumors were located within the sphenoid bone. For these cases, the dose errors were up to 6.6% and 5.4% (D98 and D95). Furthermore, for 85% of the tested patients, the mean dose to the planning target volume agreed within 2% with the calculation using the actual CT. Fictitious bone was generated in the unusual case using atlas-based methods.

CONCLUSIONS: Generally, the atlas-based method led to acceptable dose distributions. The use of common T1 sequences allows the implementation of this method in clinical routine. However, unusual patient anatomy may produce large dose calculation errors. The detection of large anatomic discrepancies using MR image subtraction can be realized, but an alternative way to produce synthetic CT numbers in these regions is still required.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app