JOURNAL ARTICLE
RESEARCH SUPPORT, N.I.H., EXTRAMURAL
Add like
Add dislike
Add to saved papers

The Impact of Direct Cardiac Output Determination On Using A Widely Available Direct Continuous Oxygen Consumption Measuring Device On The Hemodynamic Assessment of Aortic Valve.

BACKGROUND: Accurate assessment of cardiac output (CO) is essential for the hemodynamic assessment of aortic valve area (AVA). Estimation of oxygen consumption (VO2) and Thermodilution (TD) is employed in many cardiac catheterization laboratories (CCL) given the historically cumbersome nature of direct continuous VO2 measurement, the "gold standard" for this technique. A portable facemask device simplifies the direct continuous measurement of VO2, allowing for relatively rapid and continuous assessment of CO and AVA.

METHODS AND MATERIALS: Seventeen consecutive patients undergoing right heart catheterization had simultaneous determination of CO by both direct continuous and assumed VO2 and TD. Assessments were only made when a plateau of VO2 had occurred. All measurements of direct continuous and assumed VO2, as well as, TD CO were obtained in triplicate.

RESULTS: Direct continuous VO2 CO and assumed VO2 CO correlated poorly (R= 0.57; ICC =0.59). Direct continuous VO2 CO and TD CO also correlated poorly (R= 0.51; ICC=0.60). Similarly AVA derived from direct continuous VO2 correlated poorly with those of assumed VO2 (R= 0.68; ICC=0.55) and TD (R=0.66, ICC=0.60). Repeated direct continuous VO2 CO and AVA measurements were extremely correlated and reproducible [(R=0.93; ICC=0.96) and (R=0.99; ICC>0.99) respectively], suggesting that this was the most reliable measurement of CO.

CONCLUSIONS: CO calculated from direct continuous VO2 measurement varies substantially from both assumed VO2 and TD based CO, which are widely used in most CCL. These differences may significantly impact the CO and AVA measurements. Furthermore, continuous, rather than average, measurement of VO2 appears to give highly reproducible results.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app