Commissioning of the 4-D treatment delivery system for organ motion management in synchrotron-based scanning ion beams.
Physica Medica : PM 2016 December
PURPOSE: The aim of this work was the commissioning of delivery procedures for the treatment of moving targets in scanning pencil beam hadrontherapy.
METHODS: EBT3 films fixed to the Anzai Respiratory Phantom were exposed to carbon ion scanned homogeneous fields (E=332MeV/u). To evaluate the interplay effect, field size and flatness for 3 different scenarios were compared to static condition: gated irradiation or repainting alone and combination of both. Respiratory signal was provided by Anzai pressure sensor or optical tracking system (OTS). End-exhale phase and 1s gating window were chosen (2.5mm residual motion). Dose measurements were performed using a PinPoint ionization chamber inserted into the Brainlab ET Gating Phantom. A sub-set of tests was also performed using proton beams.
RESULTS: The combination of gating technique and repainting (N=5) showed excellent results (6.1% vs 4.3% flatness, identical field size and dose deviation within 1.3%). Treatment delivery time was acceptable. Dose homogeneity for gated irradiation alone was poor. Both Anzai sensor and OTS appeared suitable for providing respiratory signal. Comparisons between protons and carbon ions showed that larger beam spot sizes represent more favorable condition for minimizing motion effect.
CONCLUSION: Results of measurements performed on different phantoms showed that the combination of gating and layered repainting is suitable to treat moving targets using scanning ion beams. Abdominal compression using thermoplastic masks, together with multi-field planning approach and multi-fractionation, have also been assessed as additional strategies to mitigate the effect of patient respiration in the clinical practice.
METHODS: EBT3 films fixed to the Anzai Respiratory Phantom were exposed to carbon ion scanned homogeneous fields (E=332MeV/u). To evaluate the interplay effect, field size and flatness for 3 different scenarios were compared to static condition: gated irradiation or repainting alone and combination of both. Respiratory signal was provided by Anzai pressure sensor or optical tracking system (OTS). End-exhale phase and 1s gating window were chosen (2.5mm residual motion). Dose measurements were performed using a PinPoint ionization chamber inserted into the Brainlab ET Gating Phantom. A sub-set of tests was also performed using proton beams.
RESULTS: The combination of gating technique and repainting (N=5) showed excellent results (6.1% vs 4.3% flatness, identical field size and dose deviation within 1.3%). Treatment delivery time was acceptable. Dose homogeneity for gated irradiation alone was poor. Both Anzai sensor and OTS appeared suitable for providing respiratory signal. Comparisons between protons and carbon ions showed that larger beam spot sizes represent more favorable condition for minimizing motion effect.
CONCLUSION: Results of measurements performed on different phantoms showed that the combination of gating and layered repainting is suitable to treat moving targets using scanning ion beams. Abdominal compression using thermoplastic masks, together with multi-field planning approach and multi-fractionation, have also been assessed as additional strategies to mitigate the effect of patient respiration in the clinical practice.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app