Initial clinical experience with the Medtronic Micro Vascular Plug™ in transcatheter occlusion of PDAs in extremely premature infants.
OBJECTIVES: To describe the early multicenter, clinical experience with the Medtronic Micro Vascular Plug™ (MVP) for the occlusion of patent ductus arteriosus (PDA) in premature infants.
BACKGROUND: The MVP is a large diameter plug that can be delivered through a microcatheter for occlusion of abnormal blood vessels.
METHODS: A Retrospective review of PDA embolization procedures performed in two centers using the MVP was performed.
RESULTS: Fifteen premature infants underwent attempted PDA occlusion using the MVP. The gestational age and birth weight were 25.6 ± 2.5 weeks and 735 ± 251 g, respectively. The median weight and age at the time of the procedure were 1,210 g (700-3,500 g) and 4.5 weeks (2-12 weeks), respectively. Median procedure and fluoroscopy times were 45 and 6.5 min, respectively. The median radiation and contrast doses were 19.7 mGy and 2.4 mL/kg, respectively. Antegrade occlusion was successfully achieved in 13 patients <2 kg with only femoral venous access aided by echo guidance. The two patients >2 kg had arterial access and attempted retrograde occlusion; one of which was unsuccessful due to the PDA being short and wide. Complete closure was observed in 13 of 14 successful procedures (93%), with one patient having a small residual shunt that was not seen on follow-up. There were no complications related to the procedure or noted during follow-up (Median 11 months).
CONCLUSIONS: The MVP is a new, large-diameter vascular embolization device that may be useful for the occlusion of PDA in extremely small, premature infants. © 2016 Wiley Periodicals, Inc.
BACKGROUND: The MVP is a large diameter plug that can be delivered through a microcatheter for occlusion of abnormal blood vessels.
METHODS: A Retrospective review of PDA embolization procedures performed in two centers using the MVP was performed.
RESULTS: Fifteen premature infants underwent attempted PDA occlusion using the MVP. The gestational age and birth weight were 25.6 ± 2.5 weeks and 735 ± 251 g, respectively. The median weight and age at the time of the procedure were 1,210 g (700-3,500 g) and 4.5 weeks (2-12 weeks), respectively. Median procedure and fluoroscopy times were 45 and 6.5 min, respectively. The median radiation and contrast doses were 19.7 mGy and 2.4 mL/kg, respectively. Antegrade occlusion was successfully achieved in 13 patients <2 kg with only femoral venous access aided by echo guidance. The two patients >2 kg had arterial access and attempted retrograde occlusion; one of which was unsuccessful due to the PDA being short and wide. Complete closure was observed in 13 of 14 successful procedures (93%), with one patient having a small residual shunt that was not seen on follow-up. There were no complications related to the procedure or noted during follow-up (Median 11 months).
CONCLUSIONS: The MVP is a new, large-diameter vascular embolization device that may be useful for the occlusion of PDA in extremely small, premature infants. © 2016 Wiley Periodicals, Inc.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app