COMPARATIVE STUDY
JOURNAL ARTICLE
Performance of the PneuX System: A Bench Study Comparison With 4 Other Endotracheal Tube Cuffs.
Respiratory Care 2017 January
BACKGROUND: Cuff design affects microaspiration, a risk factor for pneumonia. We questioned whether the PneuX low-volume fold-free cuff design would prevent cuff leakage and maintain the same tracheal wall pressure as high-volume, low-pressure (HVLP) cuffs.
METHODS: We evaluated 4 HVLP-cuffed endotracheal tubes (ETTs), Hi-Lo (polyvinyl chloride [PVC]), Microcuff (polyurethane [PU]), SealGuard (PU + tapered), and TaperGuard (PVC + tapered), and the PneuX with its dedicated tracheal seal monitor. In Part 1, we determined tracheal wall pressure using each cuff's capacity to support water columns across recommended intracuff pressures. In Part 2, we evaluated the tracheal seal monitor function at recommended settings. In Part 3, we compared leakage volumes of all ETTs during 30 min of simulated mechanical ventilation or during 8 h if no leak occurred. Parts 1 and 3 were performed with/without lubrication and PEEP.
RESULTS: In Part 1, PneuX cuffs exerted an average tracheal wall pressure of 27.4 ± 2.4 cm H2 O at the recommended intracuff pressure of approximately 80 cm H2 O. Tracheal wall pressure did not differ among HVLP cuffs (19.6 ± 1.4 to 29.5 ± 1.4 cm H2 O). In Part 2, preinflation intracuff pressure affected the time to obtain tracheal seal monitor pressure attainment (P < .01). The tracheal seal monitor generated average calculated tracheal wall pressure of 33.4 ± 1.2 cm H2 O. In Part 3, PneuX ETT showed no leak across 8 h for all trials. Overall, leakage volume was lower with PU (P < .01) and PneuX (P < .01) than with PVC cuffs, regardless of shape, and lower with lubrication and/or PEEP (all P < .01). In each HVLP cuff, lubrication alone eliminated leak at an intracuff pressure of ≤30 cm H2 O.
CONCLUSIONS: The PneuX cuff generally exerted acceptable tracheal wall pressure, but the tracheal wall pressure monitor allowed pressures exceeding 30 cm H2 O in some trials and was the only ETT to prevent leak in all tests. For HVLP cuffs, leak was reduced by PU and PEEP and eliminated by lubrication.
METHODS: We evaluated 4 HVLP-cuffed endotracheal tubes (ETTs), Hi-Lo (polyvinyl chloride [PVC]), Microcuff (polyurethane [PU]), SealGuard (PU + tapered), and TaperGuard (PVC + tapered), and the PneuX with its dedicated tracheal seal monitor. In Part 1, we determined tracheal wall pressure using each cuff's capacity to support water columns across recommended intracuff pressures. In Part 2, we evaluated the tracheal seal monitor function at recommended settings. In Part 3, we compared leakage volumes of all ETTs during 30 min of simulated mechanical ventilation or during 8 h if no leak occurred. Parts 1 and 3 were performed with/without lubrication and PEEP.
RESULTS: In Part 1, PneuX cuffs exerted an average tracheal wall pressure of 27.4 ± 2.4 cm H2 O at the recommended intracuff pressure of approximately 80 cm H2 O. Tracheal wall pressure did not differ among HVLP cuffs (19.6 ± 1.4 to 29.5 ± 1.4 cm H2 O). In Part 2, preinflation intracuff pressure affected the time to obtain tracheal seal monitor pressure attainment (P < .01). The tracheal seal monitor generated average calculated tracheal wall pressure of 33.4 ± 1.2 cm H2 O. In Part 3, PneuX ETT showed no leak across 8 h for all trials. Overall, leakage volume was lower with PU (P < .01) and PneuX (P < .01) than with PVC cuffs, regardless of shape, and lower with lubrication and/or PEEP (all P < .01). In each HVLP cuff, lubrication alone eliminated leak at an intracuff pressure of ≤30 cm H2 O.
CONCLUSIONS: The PneuX cuff generally exerted acceptable tracheal wall pressure, but the tracheal wall pressure monitor allowed pressures exceeding 30 cm H2 O in some trials and was the only ETT to prevent leak in all tests. For HVLP cuffs, leak was reduced by PU and PEEP and eliminated by lubrication.
Full text links
Trending Papers
Acute and non-acute decompensation of liver cirrhosis (47/130).Liver International : Official Journal of the International Association for the Study of the Liver 2024 March 2
Guide to Utilization of the Microbiology Laboratory for Diagnosis of Infectious Diseases: 2024 Update by the Infectious Diseases Society of America (IDSA) and the American Society for Microbiology (ASM).Clinical Infectious Diseases 2024 March 6
Status epilepticus: what's new for the intensivist.Current Opinion in Critical Care 2024 Februrary 15
Administration of methylene blue in septic shock: pros and cons.Critical Care : the Official Journal of the Critical Care Forum 2024 Februrary 17
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app