PURPOSE: To report the first-in-human iliac artery experience of a new balloon-expandable covered endoprosthesis.
METHODS: A prospective, single-center pilot study recruited 30 symptomatic patients (mean age 64 years; 18 men) to evaluate the safety and early efficacy of the new Gore balloon-expandable covered endoprosthesis for the treatment of de novo or restenotic common and/or external iliac artery lesions. According to protocol, up to 2 discrete lesions could be treated with a maximum total treated length ≤110 mm. Follow-up included clinical evaluation with duplex ultrasound at 1, 6, and 12 months. Data are presented through 12-month follow-up. The primary safety endpoint was a composite of device- or procedure-related death, myocardial infarction, or amputation in the treated leg within 30 days of the index procedure. Multiple performance outcomes were also evaluated.
RESULTS: The primary 30-day safety endpoint was 0%. Per-subject estimates of primary patency, freedom from target lesion revascularization, and freedom from target vessel revascularization were 100% at 1 and 6 months and 96.6% at 12 months. Estimates of assisted primary and secondary patency were both 100% at 12 months. Freedom from major adverse events at 12 months was 100%. Most patients experienced improvements in Rutherford category, ankle-brachial index, and functional status that were sustained to 12 months.
CONCLUSION: This positive first-in-human experience with the Gore balloon-expandable covered endoprosthesis suggests this device will have an important role in the management of aortoiliac occlusive disease.
METHODS: A prospective, single-center pilot study recruited 30 symptomatic patients (mean age 64 years; 18 men) to evaluate the safety and early efficacy of the new Gore balloon-expandable covered endoprosthesis for the treatment of de novo or restenotic common and/or external iliac artery lesions. According to protocol, up to 2 discrete lesions could be treated with a maximum total treated length ≤110 mm. Follow-up included clinical evaluation with duplex ultrasound at 1, 6, and 12 months. Data are presented through 12-month follow-up. The primary safety endpoint was a composite of device- or procedure-related death, myocardial infarction, or amputation in the treated leg within 30 days of the index procedure. Multiple performance outcomes were also evaluated.
RESULTS: The primary 30-day safety endpoint was 0%. Per-subject estimates of primary patency, freedom from target lesion revascularization, and freedom from target vessel revascularization were 100% at 1 and 6 months and 96.6% at 12 months. Estimates of assisted primary and secondary patency were both 100% at 12 months. Freedom from major adverse events at 12 months was 100%. Most patients experienced improvements in Rutherford category, ankle-brachial index, and functional status that were sustained to 12 months.
CONCLUSION: This positive first-in-human experience with the Gore balloon-expandable covered endoprosthesis suggests this device will have an important role in the management of aortoiliac occlusive disease.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app