Comparison between urine culture profile and morphology classification using fluorescence parameters of the Sysmex UF-1000i urine flow cytometer.
Journal of Applied Microbiology 2017 Februrary
AIMS: To determine the usefulness of the fluorescence parameters generated by Sysmex UF-1000i flow cytometer for the rapid diagnosis of urinary tract infection by bacilli or cocci.
METHODS AND RESULTS: Urine samples (n = 1924) were studied by culture and microbiology and subsequently by cytometry, using BACT-Morph software and considering forward-scattered light (FSC) and fluorescent light scatter fluorescence parameters. BACT-Morph software showed moderate diagnostic accuracy (78·4%) to detect rod-shaped bacteria, with sensitivity of 82·4% and specificity of 62·5%. Forward-scattered (B_FSC) values of the bacterial channel were significant higher for the Gram-positive cocci category (P < 0·001). A cut-off of B_FSC ≥24·2, expressed in arbitrary units (analytical channel, ch), provided higher sensitivity (90·0%) but lower specificity (38·9%), and the diagnostic accuracy for Gram-positive cocci classification reached 62·0%.
CONCLUSIONS: Utilization of BACT-Morph software and bacterial channel fluorescence parameters (B_FSC ≥24·2 ch) offered an approximate discrimination of bacilli and cocci but the specificity was low, especially for FSC.
SIGNIFICANCE AND IMPACT OF THE STUDY: Further research is needed to establish the usefulness of flow cytometry for aetiological diagnosis.
METHODS AND RESULTS: Urine samples (n = 1924) were studied by culture and microbiology and subsequently by cytometry, using BACT-Morph software and considering forward-scattered light (FSC) and fluorescent light scatter fluorescence parameters. BACT-Morph software showed moderate diagnostic accuracy (78·4%) to detect rod-shaped bacteria, with sensitivity of 82·4% and specificity of 62·5%. Forward-scattered (B_FSC) values of the bacterial channel were significant higher for the Gram-positive cocci category (P < 0·001). A cut-off of B_FSC ≥24·2, expressed in arbitrary units (analytical channel, ch), provided higher sensitivity (90·0%) but lower specificity (38·9%), and the diagnostic accuracy for Gram-positive cocci classification reached 62·0%.
CONCLUSIONS: Utilization of BACT-Morph software and bacterial channel fluorescence parameters (B_FSC ≥24·2 ch) offered an approximate discrimination of bacilli and cocci but the specificity was low, especially for FSC.
SIGNIFICANCE AND IMPACT OF THE STUDY: Further research is needed to establish the usefulness of flow cytometry for aetiological diagnosis.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app