We have located links that may give you full text access.
Journal Article
Research Support, Non-U.S. Gov't
Single CT for attenuation correction of rest/stress cardiac SPECT perfusion imaging.
Journal of Nuclear Cardiology 2018 April
Common practice is to use separate CT scans acquired during rest and stress for attenuation correction of SPECT myocardial perfusion imaging (MPI). We evaluated using a single CT scan to correct both rest and stress SPECT scans. Studies from 154 patients were reprocessed using one CT acquired at stress to correct both rest and stress scans (1CT) and compared to correction of each scan with its own CT (2CT). Two expert readers independently read the images and determined summed stress (SSS), rest (SRS), and difference (SDS) scores. The correlation in SRS between 2CT and 1CT was r ≥ 0.88. The concordance in SDS was ≥0.84 (kappa ≥ 0.62). The mean SDS difference between 2CT and 1CT for the averaged observer was not significantly different from zero (p > 0.31). 1CT images had a small but significant increase in SRS and an increase in SDS variability. However, the mean SDS difference was similar to the mean inter-observer SDS difference for the 2CT approach (-0.08 vs -0.23, p = 0.46) and had less uncertainty (1.02 vs 2.05, p < 0.001). Thus, the differences between 1CT and 2CT are unlikely to be clinically significant, and the 1CT approach is feasible for SPECT MPI.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app