Improving beamforming performance by phased synthetic aperture imaging in medical ultrasound.
Journal of Medical Ultrasonics 2017 January
PURPOSE: Ultrasound image quality is related to the receive beamformer's ability. Delay and sum (DAS), a conventional beamformer, is combined with the coherence factor (CF) technique to suppress side lobe levels. The purpose of this study is to improve these beamformer's abilities.
METHODS: It has been shown that extension of the receive aperture can improve the receive beamformer's ability in radar studies. This paper shows that the focusing quality of CF and CF+DAS in medical ultrasound can be increased by extension of the receive aperture's length in phased synthetic aperture (PSA) imaging.
RESULTS: The 3-dB width of the main lobe in the receive beam related to CF focusing decreased to 0.55 mm using the proposed PSA compared to the conventional phased array (PHA) imaging, whose FWHM is about 0.9 mm. The clutter-to-total-energy ratio (CTR) represented by R20 dB showed an improvement of 50 and 33% for CF and CF+DAS beamformers, respectively, with PSA as compared to PHA. In addition, simulation results validated the effectiveness of PSA versus PHA.
CONCLUSION: In applications where there are no important limitations on the SNR, PSA imaging is recommended as it increases the ability of the receive beamformer for better focusing.
METHODS: It has been shown that extension of the receive aperture can improve the receive beamformer's ability in radar studies. This paper shows that the focusing quality of CF and CF+DAS in medical ultrasound can be increased by extension of the receive aperture's length in phased synthetic aperture (PSA) imaging.
RESULTS: The 3-dB width of the main lobe in the receive beam related to CF focusing decreased to 0.55 mm using the proposed PSA compared to the conventional phased array (PHA) imaging, whose FWHM is about 0.9 mm. The clutter-to-total-energy ratio (CTR) represented by R20 dB showed an improvement of 50 and 33% for CF and CF+DAS beamformers, respectively, with PSA as compared to PHA. In addition, simulation results validated the effectiveness of PSA versus PHA.
CONCLUSION: In applications where there are no important limitations on the SNR, PSA imaging is recommended as it increases the ability of the receive beamformer for better focusing.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app