Effects of Siemens TT-D carbon fiber table top on beam attenuation, and build up region of 6 MV photon beam.
Reports of Practical Oncology and Radiotherapy 2017 January
AIM: This study deals with Monte Carlo simulations of the effects which the 550 TXT carbon fiber couch can have on the relevant parameters of a 6 MV clinical photon beam in three field sizes.
BACKGROUND: According to the reports issued by the International Commission on Radiation Units and Measurements (ICRU), the calculated dose across a high gradient distribution should be within 2% of the relative dose, or within 0.2 cm of the isodose curve position in the target volume. Nowadays, the use of posterior oblique beam has become a common practice. It is clear that, in radiotherapy, the presence of the couch affects the beam intensity and, as a result, the skin dose.
MATERIALS AND METHODS: Firstly, Siemens linear accelerator validation for 6 MV photon beam was performed, and satisfactory agreement between Monte Carlo and experimental data for various field sizes was observed. Secondly, the couch transmission factor for the reference field size and depth was computed, and the skin dose enhancement by the couch was assessed.
RESULTS: The largest impact of the carbon fiber couch effect was observed for the 5 × 5 cm2 field size. Such evaluation has not been reported for this couch before.
CONCLUSION: Despite providing minimal attenuation for the primary radiation, the assumption that carbon fiber couches are radiotranslucent is not valid, and the effects of couches of this type on the transmission factor, and on the skin dose should be carefully investigated for each field size and depth.
BACKGROUND: According to the reports issued by the International Commission on Radiation Units and Measurements (ICRU), the calculated dose across a high gradient distribution should be within 2% of the relative dose, or within 0.2 cm of the isodose curve position in the target volume. Nowadays, the use of posterior oblique beam has become a common practice. It is clear that, in radiotherapy, the presence of the couch affects the beam intensity and, as a result, the skin dose.
MATERIALS AND METHODS: Firstly, Siemens linear accelerator validation for 6 MV photon beam was performed, and satisfactory agreement between Monte Carlo and experimental data for various field sizes was observed. Secondly, the couch transmission factor for the reference field size and depth was computed, and the skin dose enhancement by the couch was assessed.
RESULTS: The largest impact of the carbon fiber couch effect was observed for the 5 × 5 cm2 field size. Such evaluation has not been reported for this couch before.
CONCLUSION: Despite providing minimal attenuation for the primary radiation, the assumption that carbon fiber couches are radiotranslucent is not valid, and the effects of couches of this type on the transmission factor, and on the skin dose should be carefully investigated for each field size and depth.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app