Diastolic versus systolic ankle-brachial pressure index using ultrasound imaging & automated oscillometric measurement in diabetic patients with calcified and non-calcified lower limb arteries.

BACKGROUND: Ankle-brachial pressure index-systolic (ABI-s) can be falsely elevated in the presence of calcified lower limb arteries in some diabetic patients and therefore loses its value in this cohort of patients. We aim at investigating the feasibility of using the diastolic (ABI-d) instead of ABI-s to calculate the ABI in diabetic patients with calcified limb arteries.

METHODS: A total of 51 patients were chosen from the diabetic foot clinic. Twenty six of these patients had calcified leg arteries by Duplex scan (Group A) and 25 patients did not have calcifications in their leg arteries (Group B). Twenty five healthy volunteers were enrolled in the study for group C and they were matched with other participants from group B and A in age and sex. ABI measurement was performed using "boso ABI-system 100 machine". Systolic ABI (ABI-s) and diastolic ABI (ABI-d) were calculated based on bilateral brachial and ankle oscillometric pressures. ABI is considered normal when it is ≥0.9. Repeated measures ANOVA test was used to test for comparing mean scores for ABI-s and ABI-d across the three groups. Statistical significance is considered when P < .05.

RESULTS: The mean age of all participants (±SD) was 64.30 ± 7.1 years (range, 50-82 years). ABI-s mean ± SD was 1.3 ± 0.10 (range, 1.18-1.58) in group A patients, 1.07 ± 0.05 (range, 1-1.16) in group B patients, and 1.06 ± 0.05 (range, 1-1.16) in group C volunteers. While ABI-d mean ± SD was 1.07 ± 0.05 (range, 1.1-1.17) in group A patients, 1.06 ± 0.05 (1-1.14) in group B patients, and 1.05 ± 0.04 (range, 1.01-1.14) in group C volunteers. In group A, repeated measures ANOVA test showed statistical significant difference between ABI-s and ABI-d (P < 0.001) whereas in group B & C was not (P > 0.05).

CONCLUSIONS: ABI-d may be helpful and can be used as a complementary measure instead of ABI-s in falsely elevated ABI caused by partial incompressible vessel.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app