Performance Evaluation of the MyT4 Technology for Determining ART Eligibility.

BACKGROUND: In resource-limited countries, CD4 T-cell (CD4) testing continues to be used for determining antiretroviral therapy (ART) initiation eligibility and opportunistic infection monitoring. To support expanded access to CD4 testing, simple and robust technologies are necessary. We conducted this study to evaluate the performance of a new Point-of-Care (POC) CD4 technology, the MyT4, compared to conventional laboratory CD4 testing.

METHODS: EDTA venous blood from 200 HIV-positive patients was tested in the laboratory using the MyT4 and BD FACSCalibur™.

RESULTS: The MyT4 had an r2 of 0.82 and a mean bias of 12.3 cells/μl. The MyT4 had total misclassifications of 14.7% and 8.8% when analyzed using ART eligibility thresholds of 350 and 500 cells/μl, respectively.

CONCLUSIONS: We conclude that the MyT4 performed well in classifying patients using the current ART initiation eligibility thresholds in Mozambique when compared to the conventional CD4 technology.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app