BACKGROUND: Available atrial electrograms in implantable cardioverter defibrillators (ICDs) improve arrhythmia diagnosis, allow monitoring for atrial fibrillation, and may reduce the risk of inappropriate therapies. A recently introduced ICD system using a single-lead with floating atrial electrodes provides diagnostic capability of a dual-chamber system without placing an additional lead. Data on long-term clinical performance of this system are limited.

METHODS: We retrospectively analyzed data from 35 consecutive patients implanted with Biotronik VR-T DX devices and LinoxSmart DX leads. (Biotronik, SE & Co., Berlin, Germany) RESULTS: Of 35 patients (77% male, age 52 ± 11.28 years), 32 were followed for a mean of 432 ± 197 days (range 56-765). During implantation, average preamplified and amplified sinus P-wave amplitudes were 2.61 ± 1.39 mV (range 0.9-6.8 mV) and 8.7 ± 4.51 mV (range 1.4-18 mV), respectively. Despite statistically significant variations, the amplified P-wave amplitude measurements (calculated mean values over 3 months) remained within a clinically acceptable range during follow-up (5.4-8.7 mV). R-wave amplitude and ventricular pacing threshold measurements were stable over time. A total of 13 stored arrhythmia events (three ventricular tachycardia, eight supraventricular tachycardia, two atrial fibrillation) were reviewed. All of them showed readily interpretable atrial electrograms. Eight out of 10 (80%) supraventricular events were correctly classified by the device. Three patients received inappropriate ICD therapies.

CONCLUSION: The single-lead ICD system using a floating atrial dipole provides reliable recording of atrial signals during sinus rhythm and arrhythmias. Our data suggest that the system may offer diagnostic advantages of a dual-chamber device without potential risks of an additional atrial lead.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app