Laboratory Experience with the Liaison Analyzer in the Diagnosis of Clostridium Difficile-Associated Diarrhea.
European Journal of Microbiology & Immunology 2016 September 30
BACKGROUND: Chemiluminescent or enzyme-linked fluorescent immunoassays are commonly used to diagnose Clostridium difficile-associated diarrhea.
METHODS: The LIAISON analyzer (DiaSorin, Italy) was compared to miniVIDAS (bioMérieux, France) and, furthermore, to culture of toxigenic strains. In total, 249 native stool samples were analyzed. Sensitivities, specificities, and positive and negative predictive values were investigated. Furthermore, performance under routine conditions was assessed.
RESULTS: The glutamate dehydrogenase chemiluminescent immunoassay (GDH-CLIA) assay revealed a high sensitivity and negative predictive value. The toxins A&B assays exhibited approximately the same low sensitivity and high specificity. Technical drawbacks experienced with the LIAISON analyzer in 48% of the analyses considerably delayed the time to the first diagnostic report and interfered with laboratory routine workflow.
CONCLUSION: The analytical performance of the investigated platforms should be reflected in the context of implementation into the laboratory workflow.
METHODS: The LIAISON analyzer (DiaSorin, Italy) was compared to miniVIDAS (bioMérieux, France) and, furthermore, to culture of toxigenic strains. In total, 249 native stool samples were analyzed. Sensitivities, specificities, and positive and negative predictive values were investigated. Furthermore, performance under routine conditions was assessed.
RESULTS: The glutamate dehydrogenase chemiluminescent immunoassay (GDH-CLIA) assay revealed a high sensitivity and negative predictive value. The toxins A&B assays exhibited approximately the same low sensitivity and high specificity. Technical drawbacks experienced with the LIAISON analyzer in 48% of the analyses considerably delayed the time to the first diagnostic report and interfered with laboratory routine workflow.
CONCLUSION: The analytical performance of the investigated platforms should be reflected in the context of implementation into the laboratory workflow.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
The Effect of Albumin Administration in Critically Ill Patients: A Retrospective Single-Center Analysis.Critical Care Medicine 2024 Februrary 8
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app