Suitability of the DRI Hydrocodone/Hydromorphone Immunoassay in the Clinical Environment at a Lower Cutoff: Validation With LC-MS/MS Analysis.
Therapeutic Drug Monitoring 2016 December
BACKGROUND: We evaluated the analytical performance of the DRI hydrocodone/hydromorphone assay by comparing semiquantitative values obtained by this assay with values obtained by a liquid chromatography combined with tandem mass spectrometry (LC-MS/MS) method. We also evaluated the possibility of lowering the cutoff of the DRI assay from 300 to 100 ng/mL.
METHODS: We compared semiquantitative values obtained by the DRI assay in 97 specimens with values obtained by the LC-MS/MS method including 10 specimens containing hydrocodone and/or hydromorphone concentrations between 105.0 and 145.0 ng/mL (determined by LC-MS/MS) to determine the sensitivity at 100 ng/mL. In addition, several opioids at a concentration of 5000 ng/mL were also analyzed by the DRI assay to determine its specificity.
RESULTS: We observed no false-negative result using the DRI immunoassay in 96 specimens that showed semiquantitative values at 100 ng/mL or higher. However, one specimen containing 110 ng/mL of hydrocodone was false negative with the DRI assay (semiquantitative value 88 ng/mL, below 100 ng/mL cutoff). The semiquantitative values produced by DRI showed poor correlation with values determined by the LC-MS/MS method. The sensitivity of the DRI assay at 100 ng/mL was 90%, and the assay was very specific showing minimal cross-reactivity only with oxycodone and oxymorphone.
CONCLUSIONS: DRI immunoassay for hydrocodone/hydromorphone is a cost-effective method of screening urine specimens in the clinical environment at a lower cutoff of 100 ng/mL.
METHODS: We compared semiquantitative values obtained by the DRI assay in 97 specimens with values obtained by the LC-MS/MS method including 10 specimens containing hydrocodone and/or hydromorphone concentrations between 105.0 and 145.0 ng/mL (determined by LC-MS/MS) to determine the sensitivity at 100 ng/mL. In addition, several opioids at a concentration of 5000 ng/mL were also analyzed by the DRI assay to determine its specificity.
RESULTS: We observed no false-negative result using the DRI immunoassay in 96 specimens that showed semiquantitative values at 100 ng/mL or higher. However, one specimen containing 110 ng/mL of hydrocodone was false negative with the DRI assay (semiquantitative value 88 ng/mL, below 100 ng/mL cutoff). The semiquantitative values produced by DRI showed poor correlation with values determined by the LC-MS/MS method. The sensitivity of the DRI assay at 100 ng/mL was 90%, and the assay was very specific showing minimal cross-reactivity only with oxycodone and oxymorphone.
CONCLUSIONS: DRI immunoassay for hydrocodone/hydromorphone is a cost-effective method of screening urine specimens in the clinical environment at a lower cutoff of 100 ng/mL.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app