PURPOSE: Digital EEG has brought about greater flexibility in data interpretation but has also resulted in new and unique artifacts. As digital EEG has evolved, an increase in intensive care unit monitoring has occurred, bringing more sources of artifact to light. Aliasing as a result of a combination of compressed time base and display monitor resolution can result in appearance of spurious waveforms that can potentially skew interpretation.

METHODS: A portion of a digital EEG from an intensive care unit patient acquired at a sample rate of 1,024 Hz was reviewed at a time base of 15 mm/second on a monitor with a resolution of 1,920 × 1,080.

RESULTS: At a time base of 15 mm/second, a 60-Hz artifact resulted in the appearance of a 4-Hz delta artifact that resolved when the time base was changed to a more standard 30 mm/second.

CONCLUSIONS: A software malfunction of the digital antialiasing filter for display resulted in the appearance of a novel 4-Hz artifact.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app