Influence of echocardiographic and radiographic characteristics on atrial sensing amplitude in patients with Linox Smart S DX defibrillation leads.
Cardiology Journal 2017
BACKGROUND: Single-lead for implantable cardioverter-defibrillator (ICD) with floating atrial sensing dipole is a new diagnostic tool with the potential advantage in terms of arrhythmia discrimination. We sought to determine whether right heart size and dipole position influence atrial sensing.
METHODS: Atrial sensing (AS) amplitude was measured during implantation (PP, periprocedural), predischarge (IHFU, in-hospital follow-up) and 3-6 months after the procedure (AFU, ambulatory follow-up). Results were related to atrial dipole position in the right atrium (RA) on the basis of chest X-ray examination as well as right heart dimensions at echocardiography.
RESULTS: Twenty-four patients were included into final analysis. In 14 (58.3%) patients, sensing dipole was located in regions 1 and 2 of the RA. AS amplitude was greater in regions 1 and 2 when com¬pared to other locations (3.15 vs. 1.2 mV, p = 0.045, 7.53 vs. 3.8 mV, p < 0.001 and 5.63 vs. 2.44 mV, p = 0.017 for PP measurements, IHFU and AFU, respectively). There was a significant negative correlation between AS-PP and short RA dimension (RADs) (r = -0.56, p = 0.02), AS-IHFU and RA area (RAA) (r = -0.45, p < 0.05), AS-AFU and long RA dimension (RADl) (r = -0.46; p = 0.02), AS-AFU and RADs (r = -0,48, p = 0.02), and AS-AFU and RAA (and r = -0.52, p < 0.01). There was no relationship between AS and other right heart dimensions.
CONCLUSIONS: Larger RA size and low sensing dipole location were associated with lower AS amplitude in single-lead dual chamber ICD.
METHODS: Atrial sensing (AS) amplitude was measured during implantation (PP, periprocedural), predischarge (IHFU, in-hospital follow-up) and 3-6 months after the procedure (AFU, ambulatory follow-up). Results were related to atrial dipole position in the right atrium (RA) on the basis of chest X-ray examination as well as right heart dimensions at echocardiography.
RESULTS: Twenty-four patients were included into final analysis. In 14 (58.3%) patients, sensing dipole was located in regions 1 and 2 of the RA. AS amplitude was greater in regions 1 and 2 when com¬pared to other locations (3.15 vs. 1.2 mV, p = 0.045, 7.53 vs. 3.8 mV, p < 0.001 and 5.63 vs. 2.44 mV, p = 0.017 for PP measurements, IHFU and AFU, respectively). There was a significant negative correlation between AS-PP and short RA dimension (RADs) (r = -0.56, p = 0.02), AS-IHFU and RA area (RAA) (r = -0.45, p < 0.05), AS-AFU and long RA dimension (RADl) (r = -0.46; p = 0.02), AS-AFU and RADs (r = -0,48, p = 0.02), and AS-AFU and RAA (and r = -0.52, p < 0.01). There was no relationship between AS and other right heart dimensions.
CONCLUSIONS: Larger RA size and low sensing dipole location were associated with lower AS amplitude in single-lead dual chamber ICD.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app