JOURNAL ARTICLE
MULTICENTER STUDY
RANDOMIZED CONTROLLED TRIAL
Add like
Add dislike
Add to saved papers

Impact of Implantation Technique and Plaque Morphology on Strut Embedment and Scaffold Expansion of Polylactide Bioresorbable Scaffold - Insights From ABSORB Japan Trial.

BACKGROUND: The optimal implantation technique for the bioresorbable scaffold (Absorb, Abbott Vascular) is still a matter of debate. The purpose of the present study was to evaluate the effect of implantation technique on strut embedment and scaffold expansion.Methods and Results:Strut embedment depth and scaffold expansion index assessed by optical coherence tomography (OCT) (minimum scaffold area/reference vessel area) were evaluated in the ABSORB Japan trial (OCT subgroup: 87 lesions) with respect to implantation technique using either quantitative coronary angiography (QCA) or OCT. Strut embedment was assessed at the strut level (n=667), while scaffold expansion was assessed at the lesion level (n=81). The mean embedment depth was 63±59 µm. Balloon sizing and inflation pressure had no direct effect on strut embedment. Plaque morphology affected strut embedment [nonatherosclerotic (58.9±54.3 µm), fibroatheroma (73.3±59.6 µm), fibrous plaque (59.7±51.1 µm), and fibrocalcific plaque (-3.1±61.6 µm, negative value means malapposition), P <0.001]. The balloon-artery ratio positively correlated with the expansion index. This relationship was stronger when the OCT-derived reference vessel diameter (RVD) was used as a reference for balloon selection rather than the QCA-derived one [predilatation (Pearson correlation r: QCA: 0.167 vs. OCT: 0.552), postdilatation (QCA: 0.316 vs. OCT: 0.717)].

CONCLUSIONS: Underlying plaque morphology influenced strut embedment, whereas implantation technique had no direct effect on it. Optimal balloon sizing based on OCT-derived RVD might be recommended. However, the safety of such a strategy should be investigated in a prospective trial. (Circ J 2016; 80: 2317-2326).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app