Central X-ray beam correction of radiographic acetabular cup measurement after THA: an experimental study.

PURPOSE: Accurate assessment of cup orientation on postoperative radiographs is essential for evaluating outcome after THA. However, accuracy is impeded by the deviation of the central X-ray beam in relation to the cup and the impossibility of measuring retroversion on standard pelvic radiographs.

METHOD: In an experimental trial, we built an artificial cup holder enabling the setting of different angles of anatomical anteversion and inclination. Twelve different cup orientations were investigated by three examiners. After comparing the two methods for radiographic measurement of the cup position developed by Lewinnek and Widmer, we showed how to differentiate between anteversion and retroversion in each cup position by using a second plane. To show the effect of the central beam offset on the cup, we X-rayed a defined cup position using a multidirectional central beam offset. According to Murray's definition of anteversion and inclination, we created a novel corrective procedure to balance measurement errors caused by deviation of the central beam.

RESULTS: Measurement of the 12 different cup positions with the Lewinnek's method yielded a mean deviation of [Formula: see text] (95 % CI 1.3-2.3) from the original cup anteversion. The respective deviation with the Widmer/Liaw's method was [Formula: see text] (95 % CI 2.4-4.0). In each case, retroversion could be differentiated from anteversion with a second radiograph. Because of the multidirectional central beam offset ([Formula: see text] cm) from the acetabular cup in the cup holder ([Formula: see text] anteversion and [Formula: see text] inclination), the mean absolute difference for anteversion was [Formula: see text] (range [Formula: see text] to [Formula: see text] and [Formula: see text] (range [Formula: see text] to [Formula: see text] for inclination. The application of our novel mathematical correction of the central beam offset reduced deviation to a mean difference of [Formula: see text] for anteversion and [Formula: see text] for inclination.

CONCLUSION: This novel calculation for central beam offset correction enables highly accurate measurement of the cup position.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app