Primary posterior capsulotomy in femtosecond laser-assisted cataract surgery: In vivo spectral-domain optical coherence tomography study.
Journal of Cataract and Refractive Surgery 2016 September
PURPOSE: To evaluate whether the size of Berger space is safely predictable in patients having cataract surgery using 3-dimensional spectral-domain optical coherence tomographic (3-D SD-OCT) image analysis.
SETTING: Ruhr University Eye Hospital, Bochum, Germany.
DESIGN: Prospective interventional case series.
METHODS: Eyes having routine femtosecond laser-assisted cataract surgery were included. After wound closure, the anterior eye segment was visualized using the incorporated 3-D SD-OCT. Visualization of Berger space was performed in the axial and sagittal planes.
RESULTS: The study comprised 165 consecutive eyes. In 155 eyes, 3-D SD-OCT visualized Berger space and its dimensions were analyzed. In 72% of the cases, Berger space was large enough to perform a femtosecond laser-assisted primary posterior capsulotomy. In 24 eyes (15.5%) with a minimum axial length of 25.0 mm, Berger space was 500 μm or larger.
CONCLUSION: Femtosecond laser systems incorporating real-time SD-OCT allowed direct visualization of Berger space intraoperatively, providing surgeons with the information needed to safely create posterior capsulotomies after intraocular lens implantation.
FINANCIAL DISCLOSURE: Dr. Dick is a consultant to Abbott Medical Optics, Inc. No other author has a financial or proprietary interest in any material or method mentioned.
SETTING: Ruhr University Eye Hospital, Bochum, Germany.
DESIGN: Prospective interventional case series.
METHODS: Eyes having routine femtosecond laser-assisted cataract surgery were included. After wound closure, the anterior eye segment was visualized using the incorporated 3-D SD-OCT. Visualization of Berger space was performed in the axial and sagittal planes.
RESULTS: The study comprised 165 consecutive eyes. In 155 eyes, 3-D SD-OCT visualized Berger space and its dimensions were analyzed. In 72% of the cases, Berger space was large enough to perform a femtosecond laser-assisted primary posterior capsulotomy. In 24 eyes (15.5%) with a minimum axial length of 25.0 mm, Berger space was 500 μm or larger.
CONCLUSION: Femtosecond laser systems incorporating real-time SD-OCT allowed direct visualization of Berger space intraoperatively, providing surgeons with the information needed to safely create posterior capsulotomies after intraocular lens implantation.
FINANCIAL DISCLOSURE: Dr. Dick is a consultant to Abbott Medical Optics, Inc. No other author has a financial or proprietary interest in any material or method mentioned.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app