Shear accumulation as a means for evaluating risk of thromboembolic events in novel endovascular stent graft designs.
Journal of Vascular Surgery 2017 June
OBJECTIVE: This study proposes to establish a simulation-based technique for evaluating shear accumulation in stent grafts and to use the technique to assess the performance of a novel branched stent graft system.
METHODS: Computational fluid dynamics models, with transient boundary conditions, particle injection, and rigid walls, simplifying assumptions were developed and used to evaluate the shear accumulation in various stent graft configurations with a healthy aorta as comparison.
RESULTS: Shear streamlines are presented for the various configurations. Shear accumulation was also calculated for each configuration. The number of particles with shear accumulations >3.5 Pa-s for each configuration was compared with the shear accumulation values of commercially available mechanical aortic valves from the literature.
CONCLUSIONS: The stent graft configuration with the diaphragm does have particles with shear accumulation >3.5 Pa-s. However, the percentage of particles with shear accumulation above 3.5 Pa-s is less than the two commercially available mechanical aortic valves, and more surprisingly, is smaller than in the healthy aorta.
METHODS: Computational fluid dynamics models, with transient boundary conditions, particle injection, and rigid walls, simplifying assumptions were developed and used to evaluate the shear accumulation in various stent graft configurations with a healthy aorta as comparison.
RESULTS: Shear streamlines are presented for the various configurations. Shear accumulation was also calculated for each configuration. The number of particles with shear accumulations >3.5 Pa-s for each configuration was compared with the shear accumulation values of commercially available mechanical aortic valves from the literature.
CONCLUSIONS: The stent graft configuration with the diaphragm does have particles with shear accumulation >3.5 Pa-s. However, the percentage of particles with shear accumulation above 3.5 Pa-s is less than the two commercially available mechanical aortic valves, and more surprisingly, is smaller than in the healthy aorta.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app