Standardization of the homogeneous mobility shift assay protocol for evaluation of anti-infliximab antibodies. Application of the method to Crohn's disease patients treated with infliximab.
Biochemical Pharmacology 2016 December 16
BACKGROUND: The availability of a quantitative method to measure anti-infliximab (IFX) antibodies (ATI) would facilitate the implementation of therapeutic drug monitoring in clinical decision-making. Our aim was to standardize the homogeneous mobility shift assay (HMSA) used in the measure of ATI levels.
METHODS: In this prospective longitudinal multicenter study, 50 IFX-treated Crohn's disease (CD) patients were followed up for 54weeks. During this period 360 human serum samples were analysed. Monomeric ATI levels were measured by a quantitative HMSA-method using an anti-IFX calibrator. IFX trough levels measured by ELISA were correlated with ATI levels.
RESULTS: Using HMSA and a pure anti-idiotypic monoclonal antibody specific for IFX (anti-IFX calibrator), we measured the levels of monomeric ATI generated in Crohn's disease patients treated with IFX. Anti-IFX calibrator allowed to quantify monomeric antibodies against IFX with a low limit of quantification (3nM). The threshold level of ATI in order to classify the immunogenicity of the patients was 10nM. We observed that 24% (12/50) of IFX-treated patients developed ATI (>10nM) during the observation period (54weeks). Serum concentration of ATI higher than 10nM dramatically increased the probability (OR=51.1; 95% CI: 20.4-128.0; p<0.0001) of presenting low levels of IFX (⩽1.5nM) in serum, as observed in some CD patients treated with standard doses of the drug.
CONCLUSIONS: The HMSA-method described here allows an accurate quantification of ATI concentration in international units (IU) and therefore it could be useful in the study of the relationship between ATI concentration, infliximab level and the clinical response to the drug.
METHODS: In this prospective longitudinal multicenter study, 50 IFX-treated Crohn's disease (CD) patients were followed up for 54weeks. During this period 360 human serum samples were analysed. Monomeric ATI levels were measured by a quantitative HMSA-method using an anti-IFX calibrator. IFX trough levels measured by ELISA were correlated with ATI levels.
RESULTS: Using HMSA and a pure anti-idiotypic monoclonal antibody specific for IFX (anti-IFX calibrator), we measured the levels of monomeric ATI generated in Crohn's disease patients treated with IFX. Anti-IFX calibrator allowed to quantify monomeric antibodies against IFX with a low limit of quantification (3nM). The threshold level of ATI in order to classify the immunogenicity of the patients was 10nM. We observed that 24% (12/50) of IFX-treated patients developed ATI (>10nM) during the observation period (54weeks). Serum concentration of ATI higher than 10nM dramatically increased the probability (OR=51.1; 95% CI: 20.4-128.0; p<0.0001) of presenting low levels of IFX (⩽1.5nM) in serum, as observed in some CD patients treated with standard doses of the drug.
CONCLUSIONS: The HMSA-method described here allows an accurate quantification of ATI concentration in international units (IU) and therefore it could be useful in the study of the relationship between ATI concentration, infliximab level and the clinical response to the drug.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app