Add like
Add dislike
Add to saved papers

OPTIMISATION OF PATIENT DOSE AND IMAGE QUALITY IN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY: A PHANTOM-BASED EVALUATION.

A phantom-based study is presented aiming to optimise patient dose and image quality (IQ) in endoscopic retrograde cholangiopancreatography procedures, utilising a fluoroscopy system equipped with a flat panel detector. The patient thickness was simulated with various polymethyl methacrylate slabs, whilst IQ was evaluated using the Leeds test object. The main factors evaluated were phantom thickness, distance between phantom and detector, field of view and pulse rate. For all these factors, the dosemetric indices, entrance surface air kerma (ESAK) rate and ESAK per pulse, as well as the IQ parameters, signal-to-noise ratio and high contrast spatial resolution, were measured. Based on these measurements, the figure of merit (FOM) was estimated. The FOM and ESAK rate values indicated the optimum combination of the factors evaluated which could provide adequate clinical information, assuring minimum patient dose.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app