Effects on the Distal Radioulnar Joint of Ablation of Triangular Fibrocartilage Complex Tears With Radiofrequency Energy.
Journal of Hand Surgery 2016 November
PURPOSE: This cadaver study investigated the temperature profile in the wrist joint and distal radioulnar joint (DRUJ) during radiofrequency energy (RFE) application for triangular fibrocartilage complex resection.
METHODS: An arthroscopic partial resection of the triangular fibrocartilage complex using monopolar and bipolar RFE was simulated in 14 cadaver limbs. The temperature was recorded simultaneously in the DRUJ and at 6 other anatomic locations of the wrist during RFE application.
RESULTS: The mean temperature in the DRUJ was 43.3 ± 8.2°C for the bipolar system in the ablation mode (60 W) and 30.4 ± 3.4°C for the monopolar system in the cut mode (20 W) after 30 seconds. The highest measured temperature in the DRUJ was 54.3°C for the bipolar system and 68.1°C for the monopolar system.
CONCLUSIONS: The application of RFE for debridement or resection of the triangular fibrocartilage complex in a clinical setting can induce peak temperatures that might cause damage to the cartilage of the DRUJ. Bipolar systems produce higher mean temperatures than monopolar devices.
CLINICAL RELEVANCE: RFE application increases the mean temperature in the DRUJ after 30 seconds to a level that may jeopardize cartilage tissue.
METHODS: An arthroscopic partial resection of the triangular fibrocartilage complex using monopolar and bipolar RFE was simulated in 14 cadaver limbs. The temperature was recorded simultaneously in the DRUJ and at 6 other anatomic locations of the wrist during RFE application.
RESULTS: The mean temperature in the DRUJ was 43.3 ± 8.2°C for the bipolar system in the ablation mode (60 W) and 30.4 ± 3.4°C for the monopolar system in the cut mode (20 W) after 30 seconds. The highest measured temperature in the DRUJ was 54.3°C for the bipolar system and 68.1°C for the monopolar system.
CONCLUSIONS: The application of RFE for debridement or resection of the triangular fibrocartilage complex in a clinical setting can induce peak temperatures that might cause damage to the cartilage of the DRUJ. Bipolar systems produce higher mean temperatures than monopolar devices.
CLINICAL RELEVANCE: RFE application increases the mean temperature in the DRUJ after 30 seconds to a level that may jeopardize cartilage tissue.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app