Reproducibility and repeatability of foveal avascular zone area measurements using swept-source optical coherence tomography angiography in healthy subjects.
European Journal of Ophthalmology 2017 May 12
PURPOSE: To assess the reproducibility and repeatability of foveal avascular zone (FAZ) area measurements using swept-source optical coherence tomography angiography (SS-OCTA) in healthy subjects.
METHODS: Sixty-four eyes of 64 healthy volunteers were randomly subjected to FAZ area measurements using SS-OCTA by 2 examiners in 2 different sessions.
RESULTS: The FAZ areas measured by the first and second observer were 0.269 ± 0.092 mm2 and 0.270 ± 0.090 mm2, respectively. Within subjects, the coefficients of variations were 2.44% (95% confidence interval [CI] 1.95% to 2.93%) and 2.66% (95% CI 2.00% to 3.31%) for the first and second observers, respectively. The coefficient of repeatability average measurements of FAZ area were 0.021 mm2 and 0.024 mm2. The intraclass correlation coefficient values were 0.993 (95% CI 0.989 to 0.996) and 0.991 (95% CI 0.986 to 0.995). Interobserver and intraobserver concordance correlation coefficients ranged from 0.998 (95% CI 0.997 to 0.999) to 0.999 (95% CI 0.998 to 0.999) and from 0.989 (95% CI 0.982 to 0.993) to 0.987 (95% CI 0.979 to 0.992), respectively.
CONCLUSIONS: The FAZ area measurements by means of SS-OCTA showed high reproducibility and repeatability in healthy eyes.
METHODS: Sixty-four eyes of 64 healthy volunteers were randomly subjected to FAZ area measurements using SS-OCTA by 2 examiners in 2 different sessions.
RESULTS: The FAZ areas measured by the first and second observer were 0.269 ± 0.092 mm2 and 0.270 ± 0.090 mm2, respectively. Within subjects, the coefficients of variations were 2.44% (95% confidence interval [CI] 1.95% to 2.93%) and 2.66% (95% CI 2.00% to 3.31%) for the first and second observers, respectively. The coefficient of repeatability average measurements of FAZ area were 0.021 mm2 and 0.024 mm2. The intraclass correlation coefficient values were 0.993 (95% CI 0.989 to 0.996) and 0.991 (95% CI 0.986 to 0.995). Interobserver and intraobserver concordance correlation coefficients ranged from 0.998 (95% CI 0.997 to 0.999) to 0.999 (95% CI 0.998 to 0.999) and from 0.989 (95% CI 0.982 to 0.993) to 0.987 (95% CI 0.979 to 0.992), respectively.
CONCLUSIONS: The FAZ area measurements by means of SS-OCTA showed high reproducibility and repeatability in healthy eyes.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app