We have located links that may give you full text access.
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Incidence of Modern Alumina Ceramic and Alumina Matrix Composite Femoral Head Failures in Nearly 6 Million Hip Implants.
Journal of Arthroplasty 2017 Februrary
BACKGROUND: Because of improvements in ceramic materials and manufacturing, the incidence of ceramic failures has decreased over time. Recent concerns with corrosion have contributed to an increase in ceramic ball head utilization. The purpose of this study is to report the incidence of modern alumina bearing failures from a single major ceramic manufacturer in nearly 6 million hip implants and to identify trends in the modes of failure of these implants.
METHODS: Beginning in the year 2000, CeramTec AG (Plochingen, Germany) began a comprehensive program for reporting and gathering failure data on its products. From January 1, 2000, to December 31, 2013, over 3.2 million pure alumina (PA) and 2.78 million alumina matrix composite (AMC) ceramic ball heads were implanted worldwide. During this period, there were 672 PA and 28 AMC femoral head fractures. The fractures were analyzed with respect to time to failure, head size, and implant factors.
RESULTS: The incidence of clinical fractures of modern PA femoral heads and AMC femoral heads was 1 in 5000 (0.0201%) and 1 in 100,000 (0.0010%), respectively (P < .0001). The majority of implant failures (80%) occurred within 48 months following surgery (P < .01). Fractures were usually associated with specific events such as trauma, mismatched components, and dislocations. Large-diameter PA heads were associated with a lower rate of fracture compared to smaller-diameter femoral heads (0.0316% for 28-mm heads vs 0.0080% for heads 32 mm or greater [P < .01]). Similar trends were observed with AMC heads. The neck lengths of the femoral ball heads were also a factor: a short-taper 28-mm ball head was more likely to fracture compared to other neck lengths (P < .01).
CONCLUSION: Modern PA ceramic heads are reliable with extremely low risk of fracture. The reliability is even better with AMC heads.
METHODS: Beginning in the year 2000, CeramTec AG (Plochingen, Germany) began a comprehensive program for reporting and gathering failure data on its products. From January 1, 2000, to December 31, 2013, over 3.2 million pure alumina (PA) and 2.78 million alumina matrix composite (AMC) ceramic ball heads were implanted worldwide. During this period, there were 672 PA and 28 AMC femoral head fractures. The fractures were analyzed with respect to time to failure, head size, and implant factors.
RESULTS: The incidence of clinical fractures of modern PA femoral heads and AMC femoral heads was 1 in 5000 (0.0201%) and 1 in 100,000 (0.0010%), respectively (P < .0001). The majority of implant failures (80%) occurred within 48 months following surgery (P < .01). Fractures were usually associated with specific events such as trauma, mismatched components, and dislocations. Large-diameter PA heads were associated with a lower rate of fracture compared to smaller-diameter femoral heads (0.0316% for 28-mm heads vs 0.0080% for heads 32 mm or greater [P < .01]). Similar trends were observed with AMC heads. The neck lengths of the femoral ball heads were also a factor: a short-taper 28-mm ball head was more likely to fracture compared to other neck lengths (P < .01).
CONCLUSION: Modern PA ceramic heads are reliable with extremely low risk of fracture. The reliability is even better with AMC heads.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app