Illumination profile characterization of a light device for the dosimetry of intra-pleural photodynamic therapy for mesothelioma.
Photodiagnosis and Photodynamic Therapy 2016 December
BACKGROUND: Complete and homogeneous illumination of the pleural cavity is essential to the success of photodynamic therapy (PDT) for mesothelioma, but remains a challenge. Knowing the repartition and propagation of light around the light applicator could be the first step towards optimizing dosimetry. Here we propose a characterization method of the illumination profile of a specific light device.
METHODS: The light wand, made of a cylindrical diffuser located inside an endotracheal tube, was fixed in a tank filled with dilute 0.01% intralipid. Light dosimetry was performed around the tip of the wand using two complementary methods: direct measurements of light power with an isotropic probe and measurements of light distribution characterization.
RESULTS: Dosimetry with the isotropic probe showed an ellipse-shaped illumination. An optimized effective attenuation coefficient was deduced. Combined with the spatial representation, a theoretical illumination profile was established with iso-surfaces of fluence rate, defining a gradient light dose according to the distance from the diffuser.
CONCLUSION: A theoretical illumination profile of a light device was established and could be part of an intra-operative dosimetry system to improve light delivery during intrapleural PDT.
METHODS: The light wand, made of a cylindrical diffuser located inside an endotracheal tube, was fixed in a tank filled with dilute 0.01% intralipid. Light dosimetry was performed around the tip of the wand using two complementary methods: direct measurements of light power with an isotropic probe and measurements of light distribution characterization.
RESULTS: Dosimetry with the isotropic probe showed an ellipse-shaped illumination. An optimized effective attenuation coefficient was deduced. Combined with the spatial representation, a theoretical illumination profile was established with iso-surfaces of fluence rate, defining a gradient light dose according to the distance from the diffuser.
CONCLUSION: A theoretical illumination profile of a light device was established and could be part of an intra-operative dosimetry system to improve light delivery during intrapleural PDT.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app