We have located links that may give you full text access.
Pilot study of closed-loop anaesthesia for liver transplantation.
British Journal of Anaesthesia 2016 September
BACKGROUND: Automated titration of propofol and remifentanil guided by the bispectral index (BIS) has been used for numerous surgical procedures. Orthotopic liver transplantation (OLT) uniquely combines major changes in circulating volume, an anhepatic phase, and ischaemia-reperfusion syndrome. We assessed the behaviour of this automated controller during OLT.
METHODS: Adult patients undergoing OLT were included in this pilot study. Consumption of propofol and remifentanil was calculated for each surgery period (dissection, anhepatic, and liver reperfusion phases). Arterial blood samples were collected at several time points to allow comparison of actual with calculated propofol and remifentanil concentrations. Data are presented as median [25th and 75th percentiles] or percentage (95% confidence interval).
RESULTS: Thirteen patients were studied. System performance, defined as the percentage of time with BIS in the range 40-60, was 88% (86-94) of the total duration of anaesthesia. Propofol requirement was decreased during the anhepatic phase compared with the dissection phase (2.9 [1.9-5.0] mg kg(-1) h(-1) and 4.6 [3.5-8.1] mg kg(-1) h(-1); P<0.03) while remifentanil consumption was unchanged (0.11 [0.09-0.19] µg kg- (1) min(-1)). Bland-Altman analysis showed a weak concordance for propofol (bias of 0.7 µg ml(-1) and limits of agreement of -2.2 to +3.7 µg ml(-1)) and remifentanil (bias of 1.3 ng ml(-1) and limits of agreement -4.3 to +6.8 ng ml(-1)). No adverse events were reported during anaesthesia.
CONCLUSIONS: This pilot study indicates that automated titration of propofol and remifentanil guided by the BIS is feasible during OLT.
METHODS: Adult patients undergoing OLT were included in this pilot study. Consumption of propofol and remifentanil was calculated for each surgery period (dissection, anhepatic, and liver reperfusion phases). Arterial blood samples were collected at several time points to allow comparison of actual with calculated propofol and remifentanil concentrations. Data are presented as median [25th and 75th percentiles] or percentage (95% confidence interval).
RESULTS: Thirteen patients were studied. System performance, defined as the percentage of time with BIS in the range 40-60, was 88% (86-94) of the total duration of anaesthesia. Propofol requirement was decreased during the anhepatic phase compared with the dissection phase (2.9 [1.9-5.0] mg kg(-1) h(-1) and 4.6 [3.5-8.1] mg kg(-1) h(-1); P<0.03) while remifentanil consumption was unchanged (0.11 [0.09-0.19] µg kg- (1) min(-1)). Bland-Altman analysis showed a weak concordance for propofol (bias of 0.7 µg ml(-1) and limits of agreement of -2.2 to +3.7 µg ml(-1)) and remifentanil (bias of 1.3 ng ml(-1) and limits of agreement -4.3 to +6.8 ng ml(-1)). No adverse events were reported during anaesthesia.
CONCLUSIONS: This pilot study indicates that automated titration of propofol and remifentanil guided by the BIS is feasible during OLT.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app