Diagnostic accuracy of a pocket screening spirometer in diagnosing chronic obstructive pulmonary disease in general practice: a cross sectional validation study using tertiary care as a reference.

BMC Family Practice 2016 August 20
BACKGROUND: COPD-6™ is a lung function testing device for a rapid pre-spirometry testing to screen-out at-risk individuals not having COPD and indicating those at risk. The aim of this study was to validate COPD-6™ lung function testing (index test) in general practice in discriminating patients with COPD out of the population at risk - smokers/ex-smokers with no previous diagnosis of COPD, using measurements at tertiary care as reference standard.

METHODS: Consecutive 227 subjects (115 women, 185 smokers/42 ex-smokers, ≥20 pack-years) with no previous diagnosis of COPD, aged 52.5 (SD 6.8) years from 26 general practitioners (GPs) were recruited, lung function tested with COPD-6™, referred to the tertiary institution for repeated COPD-6™ testing followed by spirometry with a bronchodilator (salbutamol), examination, and pulmonologist consultation for the diagnosis and severity of COPD.

RESULTS: COPD was diagnosed in 43 subjects (18.9 %), with an AUC of 0.827 (95 % CI 0.769-0.875, P < 0.001) for the diagnosis of COPD when lung function was measured using COPD-6™ in GP's office with a specificity of 100 % (95 % CI, 97.95-100 %) but a very low sensitivity of 32.56 % (95 % CI, 20.49-47.48 %). Significant agreement for forced expiratory volume in 1 s measured at GP's office and at lung function lab was found (mean difference 0.01 L, p = 0.667) but not for other measured parameters (p < 0.001 for all).

CONCLUSIONS: Our study results point out that active case finding in a population at risk for COPD should be instituted (almost 20 % of undiagnosed COPD). Based on our results lung function testing with COPD-6™ can substitute spirometry testing in cases where it is not readily available to the patient/physician taken into account that the traditional FEV1/FEV6 cutoff value of <0.7 is not the only criterion for diagnosis and/or further referral.

TRIAL REGISTRATION: ClinicalTrials.gov Identifier NCT01550679 Registered 28 September 2014, retrospectively registered.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app