Journal Article
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry.

PURPOSE: To investigate agreement between the IOLMaster 700 based on swept-source optical coherence tomography (SS-OCT) and the Lenstar LS 900 based on optical low-coherence reflectometry (OLCR).

SETTING: Private practice, Castrop-Rouxelle, Germany.

DESIGN: Prospective evaluation of diagnostic test.

METHODS: Axial length (AL), keratometry (K), anterior chamber depth (corneal epithelium to lens) (ACD), lens thickness, and central corneal thickness (CCT) were measured in 183 eyes of 183 patients. Analyses used a paired t test, Pearson correlation coefficient (r), and Bland-Altman plots. Intraocular lens (IOL) power calculations were compared using the Hoffer Q, Holladay 1, and SRK/T formulas.

RESULTS: The difference in the mean AL between SS-OCT and OLCR was statistically significant but clinically insignificant (23.61 mm ± 1.27 [SD] and 23.60 ± 1.27 mm, respectively; P < .0001); the agreement and correlation were excellent. The mean K with OLCR was flatter by 0.02 diopter (D) than the mean K with SS-OCT, 43.82 ± 1.43 diopters (D) and 43.84 ± 1.43 D, respectively; the difference was not statistically significant. The mean ACD with SS-OCT was deeper by 0.03 mm than the mean ACD with OLCR, 3.22 ± 0.44 mm and 3.19 ± 0.44 mm, respectively; the difference was statistically significant (P < .001) but clinically insignificant. The differences in the mean lens thickness (OLCR 4.63 ± 0.44 mm and SS-OCT 4.59 ± 0.43 mm) and the mean CCT (OLCR 559 ± 37 μm and SS-OCT 554 ± 36 μm) were not statistically significant. There was no statistically significant difference between the median absolute errors of the 3 formulas.

CONCLUSIONS: Agreement between SS-OCT and OLCR was very good. The clinically insignificant but statistically significant differences in AL, ACD, and lens thickness did not reflect a statistically significant difference in IOL power calculation using 3 third-generation formulas.

FINANCIAL DISCLOSURE: Proprietary or commercial disclosures are listed after the references.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app