Intra- and interobserver reliability of lens equatorial length measurement using 35-MHz ultrasound biomicroscopy in dogs with cataract.
Veterinary Ophthalmology 2017 July
PURPOSE: To determine intra- and interobserver reliability of lens equatorial length measurement using 35-MHz ultrasound biomicroscopy in dogs with cataract.
METHODS: Ocular ultrasonography was performed on 28 dogs (50 eyes) before phacoemulsification. Lens equatorial length was measured on still images obtained at 35 MHz in all dogs. Each observer (A and B) obtained the measurements independently. Coefficients of variation (CVs) were used to assess intraobserver repeatability and interobserver reproducibility. Intraobserver repeatability was assessed by comparing measurements obtained on the same image on four different occasions by observer A. Interobserver reproducibility was assessed by comparing measurements obtained independently by both observers on the same image.
RESULTS: CV used to assess intraobserver repeatability was 2.4% with good agreement and no significant difference between repetitions (P > 0.05). CV used to assess interobserver reproducibility was 5.8% without significant difference between measurements obtained by the two observers (P > 0.05).
CONCLUSIONS: Intraobserver repeatability and interobserver reproducibility were considered good for lens equatorial length measurement at 35 MHz in dogs with cataract. Ultrasound biomicroscopy could be an acceptable method to assess lens equatorial dimension before phacoemulsification and intraocular lens implantation in dogs with cataract.
METHODS: Ocular ultrasonography was performed on 28 dogs (50 eyes) before phacoemulsification. Lens equatorial length was measured on still images obtained at 35 MHz in all dogs. Each observer (A and B) obtained the measurements independently. Coefficients of variation (CVs) were used to assess intraobserver repeatability and interobserver reproducibility. Intraobserver repeatability was assessed by comparing measurements obtained on the same image on four different occasions by observer A. Interobserver reproducibility was assessed by comparing measurements obtained independently by both observers on the same image.
RESULTS: CV used to assess intraobserver repeatability was 2.4% with good agreement and no significant difference between repetitions (P > 0.05). CV used to assess interobserver reproducibility was 5.8% without significant difference between measurements obtained by the two observers (P > 0.05).
CONCLUSIONS: Intraobserver repeatability and interobserver reproducibility were considered good for lens equatorial length measurement at 35 MHz in dogs with cataract. Ultrasound biomicroscopy could be an acceptable method to assess lens equatorial dimension before phacoemulsification and intraocular lens implantation in dogs with cataract.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app