Single rescuer ventilation using a bag-valve mask with internal handle: a randomized crossover trial.
American Journal of Emergency Medicine 2016 October
OBJECTIVE: To compare tidal volume received during single rescuer ventilation with a modified bag-valve mask (BVM) with integrated internal handle vs standard BVM among healthy volunteers using a manikin model.
METHODS: This study was a randomized crossover trial of adult healthcare provider volunteers performing ventilation on a manikin. We randomized participants to perform single rescuer ventilation first using either a modified BVM with integrated internal handle or a standard unmodified BVM. Participants were responsible for mask placement and delivery of 10 breaths per minute for 3 minutes as guided by a metronome. After a 3-minute rest period, they performed ventilation using the alternative device. The primary outcome measure was mean received tidal volume as measured by the manikin (IngMar RespiTrainer model). Secondary outcomes included participant reported device preference.
RESULTS: Of 70 recruited participants, all completed the study. Mean received tidal volume was higher using the modified BVM with integrated internal handle vs standard BVM by 90 mL (95% confidence interval, 60-120 mL; P< .0001). There were no significant differences in mean received tidal volume based on the order of study arm allocation. Eighty percent of participants reported preference for the modified BVM over the standard BVM (95% confidence interval, 70.0%-80.0%).
CONCLUSIONS: The modified BVM with integrated internal handle results in greater mean received tidal volume compared with standard BVM during single rescuer ventilation in a manikin model. This modified BVM design may prove a useful airway adjunct for ventilation.
METHODS: This study was a randomized crossover trial of adult healthcare provider volunteers performing ventilation on a manikin. We randomized participants to perform single rescuer ventilation first using either a modified BVM with integrated internal handle or a standard unmodified BVM. Participants were responsible for mask placement and delivery of 10 breaths per minute for 3 minutes as guided by a metronome. After a 3-minute rest period, they performed ventilation using the alternative device. The primary outcome measure was mean received tidal volume as measured by the manikin (IngMar RespiTrainer model). Secondary outcomes included participant reported device preference.
RESULTS: Of 70 recruited participants, all completed the study. Mean received tidal volume was higher using the modified BVM with integrated internal handle vs standard BVM by 90 mL (95% confidence interval, 60-120 mL; P< .0001). There were no significant differences in mean received tidal volume based on the order of study arm allocation. Eighty percent of participants reported preference for the modified BVM over the standard BVM (95% confidence interval, 70.0%-80.0%).
CONCLUSIONS: The modified BVM with integrated internal handle results in greater mean received tidal volume compared with standard BVM during single rescuer ventilation in a manikin model. This modified BVM design may prove a useful airway adjunct for ventilation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app