Add like
Add dislike
Add to saved papers

Evaluation of Anterior Segment Parameter Changes Using the Sirius after Uneventful Phacoemulsification.

PURPOSE: To investigate changes in anterior chamber depth (ACD), corneal volume (CV), anterior chamber angle (ACA), anterior chamber volume (ACV), central corneal thickness (CCT), horizontal visible iris diameter (HVID), pupil diameter (PD), and intraocular pressure (IOP) after uneventful phacoemulsification cataract surgery with intraocular lens implantation.

METHODS: A total of 132 eyes of 132 patients (87 men and 45 women) that underwent uneventful phacoemulsification cataract surgery and intraocular lens implantation were prospectively studied. The mean age of the patients was 63.68 ± 12.51 years. All patients were evaluated preoperatively and at 1 month postoperatively with the Sirius 3D Rotating Scheimpflug camera topography system. The ACD, CV, ACA, ACV, CCT, HVID, and PD measurements were recorded. IOP was measured using the Goldmann applanation tonometer, which was corrected for CCT of the Sirius device using Ehlers' formula.

RESULTS: The preoperative mean ACD, ACV, ACA, CCT, CV, PD, HVID, and IOP were 2.79 ± 0.45 mm, 124.73 ± 25.72 mm(3), 42.09 ± 7.49(0), 523.87 ± 41.97 microns, 55.37 ± 4.89 mm(3), 3.98 ± 1.23 mm, 11.72 ± 0.67 mm, and 14.74 ± 2.59 mmHg, respectively. Three months postoperatively, the mean ACD, ACV, ACA, CCT, CV, PD, HVID, and IOP were 3.45 ± 0.6 mm, 162.52 ± 23.79 mm(3), 51.46 ± 5.63(0), 526.21 ± 44.45 microns, 56.23 ± 5.12 mm(3), 2.87 ± 0.45 mm, 11.91 ± 0.75 mm, and 12.02 ± 1.83 mmHg, respectively. There was a statistically significant increase in mean postoperative ACD, ACV, ACA, CV, and HVID compared with the corresponding preoperative values (p < 0.05). CCT remained stable after surgery. Postoperative PD and IOP were significantly decreased compared to corresponding preoperative values (p < 0.05).

CONCLUSIONS: Preoperative measurements by the Sirius 3D Rotating Scheimpflug camera topography system might help surgeons to predict postoperative changes resulting from phacoemulsification and intraocular lens implantation. This is a noncontact, noninvasive, and comfortable system for patients that is highly reliable and repeatable for anterior segment measurements.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app