Journal Article
Multicenter Study
Add like
Add dislike
Add to saved papers

Evaluating the quality of implantation of percutaneous ventricular restoration device (Parachute®) by cardiac computed tomography.

BACKGROUND: The Parachute is a novel percutaneously implanted ventricular partitioning device (VPD) that has emerged as a safe and feasible treatment option for patients with heart failure following anterior wall myocardial infarction. VPD efficacy is likely dependent on optimal device placement, but to date there are no published data examining the effect of device positioning on patient outcomes.

METHODS AND RESULTS: We retrospectively identified 32 patients successfully implanted with the Parachute device, all of whom underwent cardiac computed tomography (CCT) at baseline and after 6 months of follow-up. Patients were divided into two groups based on self-reported improvement in New York Heart Association (NYHA) functional class: "not improved NYHA" (n = 12) and "improved NYHA" (n = 20). There were significant differences between both groups with regard to device positioning on follow-up CCT. Compared to patients with "improved NYHA," patients with "not improved NYHA" had longer distances from device foot to left ventricular apex (8.0 ± 4.9 vs. 2.9 ± 4.6 mm; P = 0.01), and higher lateral angles (18.0 ± 14 vs. 9.1 ± 6.8 degrees; P = 0.02), respectively. There was no significant difference between the two groups in landing zone (45.4 ± 7. vs. 45.1 ± 6.9 mm; P = 0.92) and inferior angle (14.0 ± 11.9 vs. 14.3 ± 10.1 degrees; P = 0.95). There was a numerically larger malapposition area in the "not improved NYHA" group (5.1 ± 4.5 vs. 3.2 ± 2.2 cm2; P = 0.12).

CONCLUSION: Quality of Parachute implant impacted clinical outcome, these findings should be applied prospectively in helping operators to achieve optimal implant. © 2016 Wiley Periodicals, Inc.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app