Transcatheter heart valve crimping and the protecting effects of a polyester cuff.
Morphologie : Bulletin de L'Association des Anatomistes 2016 December
INTRODUCTION: Prior to deployment, the percutaneous heart valves must be crimped and loaded into sheaths of diameters that can be as low as 6mm for a 23mm diameter valve. However, as the valve leaflets are fragile, any damage caused during this crimping process may contribute to reducing its long-term durability in vivo.
MATERIAL AND METHOD: Bovine pericardium percutaneous valves were manufactured as follows. The leaflets were sutured on a nitinol frame. A polyester cuff fabric served as a buffer between the pericardium and the stent. Two valves were crimped and one valve was used as control. The valves were examined in gross observation and micro-CT scan and then the leaflets were processed for histology and analyzed in scanning electron microscopy, light microscopy and transmission electron microscopy.
RESULT: Crimping of the valves resulted in the increase thickness of the leaflets and there was no evidence of additional delamination. The heavy prints of the stents were irregularly distributed on the outflow surface in the crimped devices and were shallow and did not penetrate throughout the thickness of the leaflets. However, the wavy microscopy of collagen fiber bundles was well preserved. They were found to remain individualized without any agglutination as shown by the regular banding appearance.
CONCLUSION: Crimping of self-deployable valves per se caused only minor damages to the leaflets. However, the procedure could be refined in order to minimize areas of high pressure and swelling of the tissue that can be accompanied with flow surface disruption and increase of the hydraulic conductance. The incorporation of a polyester buffer serves to prevent the deleterious effects that may be caused if the pericardium tissue were in direct contact with the nitinol stent.
MATERIAL AND METHOD: Bovine pericardium percutaneous valves were manufactured as follows. The leaflets were sutured on a nitinol frame. A polyester cuff fabric served as a buffer between the pericardium and the stent. Two valves were crimped and one valve was used as control. The valves were examined in gross observation and micro-CT scan and then the leaflets were processed for histology and analyzed in scanning electron microscopy, light microscopy and transmission electron microscopy.
RESULT: Crimping of the valves resulted in the increase thickness of the leaflets and there was no evidence of additional delamination. The heavy prints of the stents were irregularly distributed on the outflow surface in the crimped devices and were shallow and did not penetrate throughout the thickness of the leaflets. However, the wavy microscopy of collagen fiber bundles was well preserved. They were found to remain individualized without any agglutination as shown by the regular banding appearance.
CONCLUSION: Crimping of self-deployable valves per se caused only minor damages to the leaflets. However, the procedure could be refined in order to minimize areas of high pressure and swelling of the tissue that can be accompanied with flow surface disruption and increase of the hydraulic conductance. The incorporation of a polyester buffer serves to prevent the deleterious effects that may be caused if the pericardium tissue were in direct contact with the nitinol stent.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app