Risk of Implant Loosening After Cyclic Loading of Fusionless Growth Modulation Techniques: Nitinol Staples Versus Flexible Tether.
Spine 2017 April 2
STUDY DESIGN: Biomechanical evaluation using porcine spines.
OBJECTIVE: Compare the fixation strength of two currently used fusionless adolescent idiopathic scoliosis correction techniques following cyclic loading using porcine spines.
SUMMARY OF BACKGROUND DATA: The ability of fusionless implants to control or correct scoliosis in a growing patient requires such implants to maintain spinal fixation. Because they cross the disc, motion may weaken fixation over time.
METHODS: Eight pig spines were divided into cycled segments (T10-T13) and uncycled segments (T7-T8, L2-L3). Initial range of motion (ROM) was determined in torsion, flexion-extension, and lateral bending (0.5°/s to 1.75 N·m).Staple group (n = 4): 6 mm parallel staples were inserted on the right anterolateral spine across each intervertebral disc. Cycled segments received six staples (three adjacent discs) and uncycled segments received four staples (two separate discs).Tether group (n = 4): 5.35 × 35 mm right anterolateral vertebral body screws were placed into each vertebra. Cycled segments received four screws and uncycled segments received four screws. Screws in cycled segments were connected with a flexible tether tensioned to straight alignment.ROM of instrumented cycled segments was measured, and then segments were loaded to the measured ROM in flexion-extension (2000 cycles), lateral bending (1000 cycles), and axial rotation (2000 cycles). Implants were axially loaded to failure. Parametric tests compared pre- to postimplant ROM; nonparametric tests compared staple to screw pullout strength; P < 0.05 was significant.
RESULTS: There were no differences in ROM before instrumentation between groups. ROM was not changed except tethers decreased left lateral bending (-6.2°). Although staple pullout was less than screw pullout for cycled and uncycled segments (P < 0.05 and P = 0.057, respectively), there was no difference in pullout strength with and without cyclic loading for either group (P = 0.4).
CONCLUSION: Tethers decreased lateral bending away from the tether. Screws had almost five times greater load to failure than staples. Five thousand cycles did not result in loosening of either staple or tether screws.
LEVEL OF EVIDENCE: N/A.
OBJECTIVE: Compare the fixation strength of two currently used fusionless adolescent idiopathic scoliosis correction techniques following cyclic loading using porcine spines.
SUMMARY OF BACKGROUND DATA: The ability of fusionless implants to control or correct scoliosis in a growing patient requires such implants to maintain spinal fixation. Because they cross the disc, motion may weaken fixation over time.
METHODS: Eight pig spines were divided into cycled segments (T10-T13) and uncycled segments (T7-T8, L2-L3). Initial range of motion (ROM) was determined in torsion, flexion-extension, and lateral bending (0.5°/s to 1.75 N·m).Staple group (n = 4): 6 mm parallel staples were inserted on the right anterolateral spine across each intervertebral disc. Cycled segments received six staples (three adjacent discs) and uncycled segments received four staples (two separate discs).Tether group (n = 4): 5.35 × 35 mm right anterolateral vertebral body screws were placed into each vertebra. Cycled segments received four screws and uncycled segments received four screws. Screws in cycled segments were connected with a flexible tether tensioned to straight alignment.ROM of instrumented cycled segments was measured, and then segments were loaded to the measured ROM in flexion-extension (2000 cycles), lateral bending (1000 cycles), and axial rotation (2000 cycles). Implants were axially loaded to failure. Parametric tests compared pre- to postimplant ROM; nonparametric tests compared staple to screw pullout strength; P < 0.05 was significant.
RESULTS: There were no differences in ROM before instrumentation between groups. ROM was not changed except tethers decreased left lateral bending (-6.2°). Although staple pullout was less than screw pullout for cycled and uncycled segments (P < 0.05 and P = 0.057, respectively), there was no difference in pullout strength with and without cyclic loading for either group (P = 0.4).
CONCLUSION: Tethers decreased lateral bending away from the tether. Screws had almost five times greater load to failure than staples. Five thousand cycles did not result in loosening of either staple or tether screws.
LEVEL OF EVIDENCE: N/A.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app