Detection of 2-hydroxyglutarate in brain tumors by triple-refocusing MR spectroscopy at 3T in vivo.
Magnetic Resonance in Medicine 2017 July
PURPOSE: To test the efficacy of triple-refocusing MR spectroscopy (MRS) for improved detection of 2-hydroxyglutarate (2HG) in brain tumors at 3T in vivo.
METHODS: The triple-refocusing sequence parameters were tailored at 3T, with density-matrix simulations and phantom validation, for enhancing the 2HG 2.25-ppm signal selectivity with respect to the adjacent resonances of glutamate (Glu), glutamine (Gln), and gamma-aminobutyric acid (GABA). In vivo MRS data were acquired from 15 glioma patients and analyzed with LCModel using calculated basis spectra. Metabolites were quantified with reference to water.
RESULTS: A triple-refocusing sequence (echo time = 137 ms) was obtained for 2HG detection. The 2HG 2.25-ppm signal was large and narrow while the Glu and Gln signals between 2.2 and 2.3 ppm were minimal. The optimized triple refocusing offered improved separation of 2HG from Glu, Gln and GABA when compared with published MRS methods. 2HG was detected in all 15 patients, the estimated 2HG concentrations ranging from 2.4 to 15.0 mM, with Cramer-Rao lower bounds of 2%-11%. The 2HG estimates did not show significant correlation with total choline.
CONCLUSION: The optimized triple refocusing provides excellent 2HG signal discrimination from adjacent resonances and may confer reliable in vivo measurement of 2HG at relatively low concentrations. Magn Reson Med 78:40-48, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
METHODS: The triple-refocusing sequence parameters were tailored at 3T, with density-matrix simulations and phantom validation, for enhancing the 2HG 2.25-ppm signal selectivity with respect to the adjacent resonances of glutamate (Glu), glutamine (Gln), and gamma-aminobutyric acid (GABA). In vivo MRS data were acquired from 15 glioma patients and analyzed with LCModel using calculated basis spectra. Metabolites were quantified with reference to water.
RESULTS: A triple-refocusing sequence (echo time = 137 ms) was obtained for 2HG detection. The 2HG 2.25-ppm signal was large and narrow while the Glu and Gln signals between 2.2 and 2.3 ppm were minimal. The optimized triple refocusing offered improved separation of 2HG from Glu, Gln and GABA when compared with published MRS methods. 2HG was detected in all 15 patients, the estimated 2HG concentrations ranging from 2.4 to 15.0 mM, with Cramer-Rao lower bounds of 2%-11%. The 2HG estimates did not show significant correlation with total choline.
CONCLUSION: The optimized triple refocusing provides excellent 2HG signal discrimination from adjacent resonances and may confer reliable in vivo measurement of 2HG at relatively low concentrations. Magn Reson Med 78:40-48, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app