Identifying a Safe Range of Stimulation Current for Intraoperative Neuromonitoring of the Recurrent Laryngeal Nerve: Results from a Canine Model.

BACKGROUND: Intraoperative neuromonitoring (IONM) of the recurrent laryngeal nerve (RLN) has been widely applied during thyroid surgery. However, the safe range of stimulation intensity for IONM remains undetermined.

METHODS: Total thyroidectomies were performed on twenty dogs, and their RLNs were stimulated with a current of 5-20 mA (step-wise in 5 mA increments) for 1 min. The evoked electromyography (EMG) of vocal muscles before and after supramaximal stimulation were recorded and compared. Acute microstructural morphological changes in the RLNs were observed immediately postoperatively under an electron microscope.

RESULTS: The average stimulating threshold for RLNs stimulated with 15 mA and 20 mA showed no significant changes compared to the unstimulated RLNs (15 mA group: 0.320 ± 0.123 mA vs. 0.315 ± 0.097 mA, P = 0.847; 20 mA group: 0.305 ± 0.101 mA vs. 0.300 ± 0.103 mA, P = 0.758). Similar outcomes were shown in average evoked EMG amplitude (15 mA group: 1,026 ± 268 μV vs. 1,021 ± 273 μV, P = 0.834; 20 mA group: 1,162 ± 275 μV vs. 1,200 ± 258 μV, P = 0.148). However, obvious acute microstructural morphological changes were observed in the nerves that were stimulated with 20 mA.

CONCLUSIONS: A stimulation intensity less than 15 mA might be safe for IONM of the RLN.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app